Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
40
Default settings
Network configuration is done ex factory as indicated on the serial number label or in the technical specifications. The table
below shows the supported configurations for the supported interface protocols (default settings are printed in boldface):
Protocol
ProPar
(RS232)
FLOW-BUS
(RS485)
Modbus
(RTU/ASCII)
PROFIBUS DP
DeviceNet™
Bus Address
3
3
…125
1
…247
0…
126
0…
63
Baud Rate
9600
19200
38400
57600
115200
230400
460800
187500
400000
9600
19200
38400
56000
57600
115200
128000
256000
(autodetect)
9600
19200
45450
93750
187500
500000
1500000
3000000
6000000
12000000
125000
250000
500000
Parity
0
0
0, 1,
2
2
0
Network configuration for EtherCAT® and PROFINET is done automatically via the Ethernet protocol.
4.7
Special parameters
Init Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
82/64
7
0/10
0x000A/11
Init Reset
is used to unlock secured parameters (designated by a
symbol) for writing. This parameter can be set to the
following values:
Value
Description
64
unlocked, secured parameters can be read and written to
82
locked, secured parameters are read-only
At power-up,
Init Reset
is always set to ‘Locked’ (value 82).
Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
R
0…7
114
115/8
0x0E68/3689
This parameter is used to reset the program, counter or alarms.
Value
Description
0
No reset
1
Reset counter
2
Reset alarm
3
Restart batch counter
4
Reset and disable counter
5
Reset firmware program (soft reset)
6
Reset
Alarm info
error bit
7
Reset
Alarm info
warning bit
The
Reset
parameter may be disabled by
Reset Alarm Enable
or
Reset Counter Enable
. Make sure the value is accepted
by sending 0 first.