Polycold Cryochiller
Installation
Installation and Operation Manual
Inspect and Unpack the Polycold Cryochiller
Brooks Automation
214072 Revision B
4-5
1.
Carefully inspect all surfaces of the shipping containers for signs of damage or stress.
Note that there may be more than one shipping container.
2.
If the container has signs of damage:
•
Record the damage and notify the carrier immediately
.
•
Assess the damage to the shipping container with the carrier.
•
Determine if the Polycold Cryochiller is damaged.
•
If the Polycold Cryochiller is damaged, STOP THIS PROCEDURE!
Do NOT install a damaged unit.
Notify Brooks Automation of the damage to the unit.
•
Arrange with the carrier and Brooks Automation to repair the damage or replace the unit.
Refer to the Brooks Automation contact information in the Appendix of this manual.
•
If there is no damage, continue this procedure
3.
Remove strapping and fasteners that hold the cardboard enclosure to the palette.
Examine the contents for damage.
Report any damage that is discovered.
4.
Remove the lower braces from the palette that block the base frame of the Cryochiller.
High Pressure Rupture
If a high pressure circuit is ruptured, such as a refrigerant circuit,
equipment damage could occur. Cuts and eye injury from flying
objects could result in death or serious injury.
•
Inspect the refrigerant circuit and entire unit before performing
any installation procedures.
Heavy Object - Lift / Tip Over
Polycold Cryochiller refrigeration unit weighs a minimum of 533kg
(1175 lb) and up to 544kg (1200 lb). Failure to properly lift or move
this device may result in death or serious injury.
•
Use proper lifting devices and techniques when moving the Poly-
cold Cryochiller.