LEGACY_REMOTE,RMT_SIGNAL_LOGIC Legacy Remote Output Signal Logic
Purpose:
Get/Set the Legacy Remote, remote output signal logic. Default setting is to
remain compatible with remote signal functionality on PFC. The remote output
signal will be Low when the Compressor is off regardless of unit being in Local
or Remote control.
Monitor Data:
N/A
Parameter(s):
? or [n]
[n]
0 = Remote output signal will be:
Low when compressor is Off in Local Control
High when compressor is Off in Remote Control
Original MaxCool remote signal logic
1 = Remote output signal will be:
Low when compressor is Off in Local Control
Low when compressor is Off in Remote Control
Original PFC remote signal logic
Return Value:
[A] or [A][n]
Data Type:
[n]: Integer
Default Value:
[n] = 1
Min: [n] = 0
Max: [n] = 1
Units:
N/A
Security Level
Get: 1
Set: 3
Example:
“LEGACY_REMOTE,RMT_SIGNAL_LOGIC,0” Set the Legacy Remote,
remote signal logic to the original MaxCool logic.
Interface
RS-232, Ethernet