3.1 OD Adjustment (Optical Density Learning)
(1) Press the [Start] button on your computer and click "All Programs". Click and open
"Standard Learning Tool II" to start "Standard Learning Tool II".
(2) When "Please Input ID or Station Name" and "Please Input Password" windows appear,
enter "service".
*
Password will be hidden when you enter.
(3) Select "OD" and click the [Ok] button.