Chapter 5 – Maintenance and Troubleshooting
Troubleshooting
BB1224 – 11
57
Type 2239 B
Hand-Arm Vibration Functions User Manual
5.2.3
Handling the instrument
●
Do not attempt to open the instrument. There are no user-
serviceable parts inside. If you think your instrument re-
quires service, then please contact your Brüel & Kjær dealer
●
Do not allow the instrument to get wet
●
Never mix different makes or types of battery
●
Never mix charged and discharged batteries
●
Do not allow fully discharged batteries to remain inside
the instrument
●
Protect the instrument from impact. Do not drop it. Trans-
port it in the supplied shoulder bag
5.3
Troubleshooting
The wire that leads from the vibration meter to the acceler-
ometer may become damaged after extensive use, and will
thereafter provide a weak electrical connection. If your read-
ings start to seem unreliable, the first thing you should try is
to replace this wire. We recommend you keep a spare on hand,
just in case (see section 6.2 for ordering information). If this
does not help, then you may need a new calibration or there
may be something wrong with the accelerometer or the vibra-
tion meter itself. Please contact your local Brüel & Kjær dealer
for assistance.
For help with error messages or other warnings that appear
on the screen, refer to section 5.4.