Chapter 5 – Maintenance and Troubleshooting
Warnings and Error messages
BB1224 – 11
59
Type 2239 B
Hand-Arm Vibration Functions User Manual
Message
Occurs
Solution
When you try to save
a record when all 40
records are full.
Erase records or
take manual
records. See section
4.1.2 for more sug-
gestions.
On start-up if there is
a problem with the
memory or program.
Contact your
Brüel & Kjær dealer
for assistance.
Normally, as
described in section
2.5 (not an error).
However, if it appears
unexpectedly on
start-up, then your
back-up battery is
probably flat. In this
case, the clock and all
settings will also be
reset. This means that
the instrument
requires a new fac-
tory calibration.
Contact your
Brüel & Kjær dealer
to arrange for a
new factory cali-
bration.
If you press
when there are no
records saved and the
buffer is empty.
Make and store
measurements
before using this
function.
*** WARNING ****
No Free Records
to Store
Overall Results
2239 SELFTEST
Memory:....FAILED
Program:...FAILED
SET-UP
Language
> English
*** WARNING ****
No Data
Available
bb122411.book Page 59 Tuesday, December 11, 2001 2:34 PM