Operating the VIBROCONTROL 1500
VIBROCONTROL 1500
VC1500EN / C103845.002 / V 09 / 14.04.2015
- 45 -
With
L1BC, L2BC, L1temp
and
L2temp
no relay response lag time
for the
limit value violation is preselectable. Signaling of these events takes place
immediately after violation of the set limit values.
Fig. 41: Limit settings
With further paging through the menu points using the buttons
and
the
display with setting options for the measuring range and input filter frequency
are reached.
Fig. 42: Setting for the measuring range and highpass corner frequency
Possible measuring range:
10; 20; 40; 50; 100; 200
mm/s
(0.5; 1,0; 2; 2.5; 5.0; 10.0
in/s
)
A plausibility check when changing the measurement range with respect to
the pre-set limit values is not performed. This means, for example, it is
possible to set the measuring range smaller than the limit values. The correct
configuration is the sole responsibility of the user.
Possible high pass corner frequency:
1, 3, 10 Hz. Please note that through selection of the corner frequency all
vibrations with frequencies lower than the selected corner frequency will be
suppressed.
Note:
When a high pass corner frequency is edited it is necessary to carry
out a reset of the instrument. This happens automatically when the
selected changes to the settings are stored.