Chapter 2 — Operation
Circuit Adjustment for Acceleration Measurement
Preamplifiers Types 2663 and 2663 B
Brüel & Kjær
16
The value of
R
G
can be calculated as follows:
(2.5.4)
To adjust the sensitivity, exchange resistor
R
G
with a resistor of the value calculated
from formula 2.5.4.
2.5.5
Grounded/Floating
Fig.2.10
Grounded/Floating connection shown with (left) Floating and (right) Grounded input config-
uration
This has already been discussed in
and is used when choosing between
single-ended grounded and single-ended floating input connections.
R
G
1.69
10
5
Gain
--------------------------
10