System Components
20
H157655_1_008
3.2
Console and Other System Components
The next table lists the various parts of the console, monitoring & control units. Please also
refer to the floor plan diagrams beginning in the chapter
. These scaled
diagrams provide an idea of where the various pieces of NMR equipment should be placed.
1
2
3
4
5
6
Figure 3.1: Spectrometer and Magnet Control
1.
The
AVANCE console
main cabinet, where the actual NMR data acquisition is
performed.
2.
The
probe
, which is designed to hold the sample, transmit radio frequency signals
which excite the sample and receive the emitted response. The probe is inserted
into the bottom of the magnet and sits inside the room temperature shims. Coaxial
cables carry the excitation signals from the console amplifiers to the probe and
the NMR signal back from the sample to the receiver.
3.
The
HPPR/2
amplifies, filters and routes the NMR response signals from the
probe to the RX22 receiver. It switches the RF transmitter output to the probe.
4.
The
BCU-II Unit
delivers very cold gas, either nitrogen or dry air, through a
flexible isolated non-magnetic transfer line. It is possible to control the sample
temperature down to -60°C inside the probe for solid or liquid NMR applications.
5.
The
BCU-I Unit
cools VT gas to allow proper sample temperature regulation. The
unit reduces the temperature of the air input (supplied by the variable-temperature
unit) and provides cooling of the NMR sample within the magnet to at least -5 °C
for a room temperature of 25 °C.
6.
The
workstation
acts as the operational computer for the user processing NMR
data and sending/receiving data to/from the acquisition computer in the main
console.
Summary of Contents for AVANCE
Page 6: ...Contents vi H157655_1_008 ...
Page 24: ...System Components 24 H157655_1_008 ...
Page 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Page 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Page 72: ...Utility Requirements 72 H157655_1_008 ...
Page 82: ...Floor Plan 82 H157655_1_008 ...
Page 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Page 102: ...Contact 102 H157655_1_008 ...
Page 104: ...List of Figures 104 H157655_1_008 ...
Page 109: ...H157655_1_008 109 ...
Page 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...