Ceiling Height Requirements
H157655_1_008
37
Ceiling Height Requirements
The number in the following tables corresponds to a standard magnet configuration using a
minimum height stand. The ceiling height can increase up to 250 mm based on the model of
the magnet and stand. Refer to the table in Magnet Stand Transport Dimensions for
specification on the magnet model and stand.
Magnet Type
Special LHE Transfer Line
for Reduced Ceiling
Height (m)
Standard LHE Transfer
Line (m)
400/54 Ascend Aeon
2.66
3.04
500/54 Ascend Aeon
2.66
3.04
600/54 Ascend Aeon
3.00
3.39
700/54 Ascend Aeon
3.00
3.39
Table 5.2:
Minimum Ceiling Height Requirements: Standard Bore 54 mm
Magnet Type
Special LHE Transfer Line
for Reduced Ceiling
Height (m)
Standard LHE Transfer
Line (m)
400/89 Ascend Aeon
3.25
3.64
500/89 Ascend Aeon
3.25
3.64
600/89 Ascend Aeon
---
3.65
700/89 Ascend Aeon
3.40
3.79
Table 5.3:
Minimum Ceiling Height Requirements: Wide Bore 89 mm