CryoProbe and Other Accessories
H157655_1_008
89
10.2.2
Space Requirements and Specifications
10.2.2.1
Indoor Helium Compressors
The indoor helium compressors (air or water cooled) have space requirements to allow for
airflow and servicing the unit. The minimum room space needed is 1.25 m (width) x 0.8 m
(depth) x 0.7 m (height).
Sumitomo Type:
F-50L
F-50H
CSA-71A
Type of Compressor:
Water-cooled
Water-cooled
Air-cooled
Dimension:
Width: 45 cm
Depth: 48.5 cm
Height: 59.1 cm
Width: 45 cm
Depth: 48.5 cm
Height: 59.1 cm
Width: 55 cm
Depth: 55 cm
Height: 88.5 cm
Weight:
120 kg
120 kg
140 kg
Voltage:
3 x 200 V @ 60 Hz
3 x 400 V (380,
400, 415 @ 50 Hz,
or 460, 480 @ 60
Hz
3 x 200 V @ 50/60
Hz US no plug
supplied
Mainly deliverd to:
USA/CA/JP
EU/CH
USA/CA/JP
Operating Current:
26 A
13 A
25 A
Minimum Circuit Ampacity:
35 A
17 A
35 A
Maximum Fuse Size:
60 A
30 A
60 A
Compressor LRA:
160 A
75 A
100 A
Power Requirements
Minimum:
Recommended:
9 kVA
12 kVA
9 kVA
12 kVA
9 kVA
12 kVA
Power Consumpton
Cool Down max. 50/60 Hz:
Steady State max. 50/60
Hz:
7.2/8.3 kW
6.5/7.5 kW
7.2/8.3 kW
6.5/7.5 kW
7.2/8.3 kW
6.5/7.2 kW
Water supply to remove
the heat load (cooling
power > 8.3 kW):
Flow: 420 l/hour
Temperature: 4-28°C
Water Quality: PH 6.5-8.2
Hardness: mg[CaCo3]/l < 200
Molybdate-reactive silicate: < 50 mg/l
Suspended matter: < 10 mg/l
Maximum pressure: < 7 bar (100 PSI)
---
Helium Gas Supply:
A high purity helium (He) gas cylinder (6 N = 99.9999% or
better, maximum impurity concentration 1 ppm).
Acoustic Noise:
Maximum 60 dB(A), 2 meters distant.
Ambient Operating
Temperature:
5 to 28°C (41 to 82.4°F)
30°C
Table 10.3:
Technical Data for Indoor Compressors
Summary of Contents for AVANCE
Page 6: ...Contents vi H157655_1_008 ...
Page 24: ...System Components 24 H157655_1_008 ...
Page 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Page 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Page 72: ...Utility Requirements 72 H157655_1_008 ...
Page 82: ...Floor Plan 82 H157655_1_008 ...
Page 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Page 102: ...Contact 102 H157655_1_008 ...
Page 104: ...List of Figures 104 H157655_1_008 ...
Page 109: ...H157655_1_008 109 ...
Page 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...