CryoProbe and Other Accessories
H157655_1_008
95
10.6
CryoFit
This section describes the requirements for the site planning for the CryoFit.
10.6.1
Introduction
A CryoFit converts a standard CryoProbe into a flow probe within a few minutes. The CryoFit
is a long metal pole with a flow cell. The flow cell is positioned inside the CryoFit during
storage and installation. The outlet capillary is attached to the flow cell and runs inside the
metal-pole to the top of the CryoFit. The inlet capillary is attached to the flow cell and is guide
during the installation procedure through the VT-air inlet of the probe to the outside.
CryoFits are compatible with all newer 5 mm He-cooled CryoProbes as well as Prodigy
CryoProbes.
The CryoFit is inserted from the top of the magnet into the upper part of the shim system
(BST Upper Part) is pushed into the magnet until the end of the CryoFit sits in the spinner
turbine. After this installation procedure the flow cell is pushed into the NMR probe so that is
located in the area of the NMR coils like a sample tube.
10.6.2
Installation Requirements
Two essential requirements have to be observed for the CryoFit.
1. The CryoFit must be
long enough
to that it fits into the BST and the flow cell can reach
the magnet center. This requirement is automatically checked during the order review.
– If a CryoFit is ordered as accessory for an already existing, older magnet the length
should be verified as the automatic conversion might not work properly.
– The automatically selected CryoFits are will be compatible with all types of sample
changers and included some security margin. In case a problem with the available
ceiling height –see (2) below–, you can verify if the smaller CryoFit fits the current
instrument configuration, for example if no SampleChanger is included.
2. The CryoFit requires a minimum ceiling height so that the distance between BST upper
part and the ceiling allows the insertion of the CryoFit.
– The calculation of the required ceiling height depends on many parameters (type of
magnet, magnet feet, BST length, type of sample changer).
For the calculation of the magnet ceiling height (MCH) refer to the CryoFit Site Planning
Guide.
Summary of Contents for AVANCE
Page 6: ...Contents vi H157655_1_008 ...
Page 24: ...System Components 24 H157655_1_008 ...
Page 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Page 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Page 72: ...Utility Requirements 72 H157655_1_008 ...
Page 82: ...Floor Plan 82 H157655_1_008 ...
Page 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Page 102: ...Contact 102 H157655_1_008 ...
Page 104: ...List of Figures 104 H157655_1_008 ...
Page 109: ...H157655_1_008 109 ...
Page 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...