Environment and Site Survey Measurement
48
H157655_1_008
7.1.5
Floor Vibration Guidelines: Bruker Nano-C and Nano-D
ADI damping system (Air Spring Damped Isolators with vertical damping)
Frequency
Maximum Velocity in
Vertical Direction
Maximum Velocity in
Horizontal Direction
1-5 Hz
3 µm/s
1.6 µm/s
5-20 Hz
3…12 µm/s
1.6…6.4 µms
>20 Hz
12 µm/s
6.4 µm/s
Table 7.3:
ADI Damping System (Elastomeric Isolators)
Figure 7.5: Example of the Bruker Tolerance Level for ADI Dampers. Resolution <0.125 Hz, 1…200 Hz,
> 30 min, Average Measurement 1 (displayed in velocity)