Characteristics
Inlet Temperature Range [°C] (°F)
[4.0 ~ 28.0] (39.2 ~ 82.4)
Inlet Pressure Range [MPa] (psig)
[0.10 ~ 0.69] (14.5 ~ 100)
Flow Rate [liter/min.] (gallon/min.)
[4.0 ~ 10.0] (1.1 ~2.6)
Pressure Drop [MPa] (psig)
[0.025 ~ 0.085] (3.55 ~ 12.1)
Heat Output [kW ] (BTU/Hr) into
the water (residual of 0.5 kW into
air)
<Steady State>
[< 6.5] (< 22180) for 50Hz
[< 7.5] (< 25590) for 60Hz
<Maximum>
[< 7.2] (< 24570) for 50Hz
[< 8.3] (< 28320) for 60Hz
Quality
pH Value
Electrical Conductivity
Chloride Ion
Sulfate Ion
M-Alkalinity
Total Hardness
Calcium Hardness
Ionic Silica
Iron
Copper
Sulfide Ion
Ammonium ion
Residual Chlorine
Free Carbon Dioxide
Stability Index
Suspended Matter
Particle Size
6.5 to 8.2 at 25 °C (77 °F)
< 80 mS / m
< 200 mg/liter
< 200 mg/liter
< 100 mg/liter
< 200 mg/liter
< 150 mg/liter
< 50 mg/liter
< 1.0 mg/liter
< 0.3 mg/liter
None, Not detectable
< 1.0 mg/liter
< 0.3 mg/liter
< 4.0 mg/liter
6.0 to 7.0
< 10 mg/liter
< 100 μm
Table 10.2:
Cooling Water Specifications
• Ambient operating temperature 5–28ºC.
Figure 10.4: Helium Compressor - Indoor Water Cooled
Summary of Contents for AVANCE
Page 6: ...Contents vi H157655_1_008 ...
Page 24: ...System Components 24 H157655_1_008 ...
Page 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Page 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Page 72: ...Utility Requirements 72 H157655_1_008 ...
Page 82: ...Floor Plan 82 H157655_1_008 ...
Page 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Page 102: ...Contact 102 H157655_1_008 ...
Page 104: ...List of Figures 104 H157655_1_008 ...
Page 109: ...H157655_1_008 109 ...
Page 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...