Specifications
7
Common Characteristics
7.1
Table 7.1.
Amplifier Common Characteristics
Constant Internal Protection
Supplies, fans faults and over temperature.
Forward Power : peak & CW power, pulse width and duty cycle.
Reflected Power : peak & CW power, self resetting protection
shuts the amplifier off if the load VSWR is excessive.
Front Panel Indicators
Amplifier Status Led Board.
Front Panel Interfaces
1 x I/O 8 pins RJ45 connector.
Front Panel Controls
1 x AC line ON/OFF switch,
1 x SEL H500/H50 control signal.
Front Panel Connectors
3 x RF input,
2 x RF output,
1 x blanking input (gating).
Rear Panel Connectors
1 x AC line in socket.
Cooling System
Forced-air cooling (from front to rear).
Size
19" rack cabinet x 3U height x 520mm depth.
Weight
19kg
Power Requirements
208-230 VAC ± 10% single phase 50-60Hz.
Bruker Biospin part number W1304006.
Consumption max. 0.90kVA.
Summary of Contents for BLAH500 E
Page 6: ...6 51 BRUKER BIOSPIN Operating Service Manual Version 002 General Information ...
Page 10: ...10 51 BRUKER BIOSPIN Operating Service Manual Version 002 Safety ...
Page 14: ...14 51 BRUKER BIOSPIN Operating Service Manual Version 002 Installation ...
Page 28: ...28 51 BRUKER BIOSPIN Operating Service Manual Version 002 Technical Description ...
Page 44: ...44 51 BRUKER BIOSPIN Operating Service Manual Version 002 Specifications ...
Page 48: ...48 51 BRUKER BIOSPIN Operating Service Manual Version 002 Figures ...
Page 50: ...50 51 BRUKER BIOSPIN Operating Service Manual Version 002 Tables ...
Page 51: ...Operating Service Manual Version 002 BRUKER BIOSPIN 51 51 End of Document ...