46 (51)
BRUKER BIOSPIN
Operating & Service Manual Version 002
Service Information and Maintenance
Figure 8.1. Push Fan Assembly
4. Unscrew the 2 screws on the top of the Push fan assembly.
5. Remove the Push fan assembly.
6. Place correctly the new fan assembly in the bottom holes of the RF module
and screw it on the top.
7. Connect all wires (status and supply).
8. Connect line cord and turn on the amplifier. Note that the fans are turning and
no OVERHEAT status led appears on front panel.
9. Put the coverage plate on the amplifier and screw it.
10. Put the amplifier in the NMR console, connect all cables on the front panel and
the line cord on the rear panel.
Push fan supply
connector
Connector J18 on
BLA Control Board
Screws for Push
fan assembly
Fans status wires from
BLA Control board
connector J18