Error Descriptions
Daedalus Lock Manual
Spectrospin AG
99
LOCK RX AGC Voltage Failure
Error 158
The control voltage for the Receiver AGC Amplifiers is
checked by the Microprocessor.
Possible Causes:
- The Pulse Section in the transmitter is not working correctly
- 10 MHz reference is not connected.
- Problems with the serial bus
- AGC Voltage Control Unit is not working correctly (L-RX:
IC10, IC12)
- Temperature Compensation for AGC Voltage does not work
(IC25)
- The Receiver A/DC (IC9) does not work properly
Diagnostic Channel
Pulsbank Nr 10
Signalname UAGC
ADC_CONV
PL_CLK
HC_10MHz
Serial Bus
LOCK RxTyp <> TxTyp
Error 160
Lock Receiver and Transmitter have incompatible
frequencies. Incorrect Receiver or Transmitter is in the BSMS
rack. If this message is displayed after a long period of
operation the microprocessor analog input is probably defect.
LOCK RxOpt <> TxOpt
Error 161
Incompatible Options modules are installed in the Receiver
and Transmitter.
LOCK Syntax Error
Error 20
Appears if the CPU transfers undefined Syntax.
LOCK Too much Noise on ADC
Error 153
With minimal RFGain, the Receiver Noise is measured.
Possible Causes:
- The Pulse Section in the transmitter is not working correctly
- 10 MHz reference is not connected.
- Failure in Receiver Acquisition Section (IC9)
- Too much noise in Receiver HF-Section
- Too much noise in Receiver NF-Section
- Failure in AGC Control Voltage Unit
If the Error ‘Lock RX AGC Voltage Failure’ appears in
addition then attend to it first.
Diagnostic Channel
Pulsbank Nr 5
NF1..NF4
NF_OUT
UAGC
UAGC2
ADC_CONV
PL_CLK
HC_10MHz
RP1
LOCK Watchdog
Error 3
The watchdog in the lock system was activated. The users
tried to execute functions which are not supported while the
watchdog is on. Deactivate the watchdog in the BSMSTOOL
first and try it again.
LOCK Wrong Datacount
Error 31
The lock application software is a INTEL Hex file. If the
length of any line in this file violates the format specifications
downloading will stop immediately.
Your downloaded file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
Table 12. Lock Error Messages in alphabetical order
ERROR MESSAGE
ERROR DESCRIPTION
SIGNALS
Note: Signals of
the serial bus
(power, gain
phase, shift etc.)
are only active
during a process
change
Summary of Contents for DAEDALUS
Page 1: ...BRUKER BRUKER SPECTROSPIN AG SERVICE MANUAL DAEDALUS LOCK Version BSMS 004 ...
Page 6: ...Contents 6 Spectrospin AG Daedalus Lock Manual ...
Page 14: ...8 Spectrospin AG Daedalus Lock Manual ...
Page 16: ...10 Spectrospin AG Daedalus Lock Manual ...
Page 26: ...General Description 20 Spectrospin AG Daedalus Lock Manual ...
Page 36: ...Operation 30 Spectrospin AG Daedalus Lock Manual ...
Page 56: ...Lock Transmitter 50 Spectrospin AG Daedalus Lock Manual ...
Page 64: ...Lock Receiver 58 Spectrospin AG Daedalus Lock Manual ...
Page 68: ...19F RX Option 62 Spectrospin AG Daedalus Lock Manual ...
Page 85: ...Installation Daedalus Lock Manual Spectrospin AG 79 Installation 11 2 ...
Page 86: ...Z0 Compensation Option 80 Spectrospin AG Daedalus Lock Manual ...
Page 90: ...Technical Data 84 Spectrospin AG Daedalus Lock Manual ...
Page 92: ...Trouble Shooting 86 Spectrospin AG Daedalus Lock Manual ...
Page 98: ...BSMS service tool 92 Spectrospin AG Daedalus Lock Manual ...
Page 107: ...Error Descriptions Daedalus Lock Manual Spectrospin AG 101 ...
Page 108: ...Lock Error Messages 102 Spectrospin AG Daedalus Lock Manual ...
Page 114: ...Appendix 108 Spectrospin AG Daedalus Lock Manual ...
Page 122: ...116 Spectrospin AG Daedalus Lock Manual ...
Page 123: ...Daedalus Lock Manual Spectrospin AG 117 Lock Transmitter Index B ...
Page 124: ...118 Spectrospin AG Daedalus Lock Manual ...
Page 125: ...Daedalus Lock Manual Spectrospin AG 119 Lock Transmitter Index C ...
Page 126: ...120 Spectrospin AG Daedalus Lock Manual ...
Page 127: ...Daedalus Lock Manual Spectrospin AG 121 19F Transmitter Option Index A ...
Page 128: ...122 Spectrospin AG Daedalus Lock Manual ...
Page 129: ...Daedalus Lock Manual Spectrospin AG 123 Lock Receiver Index C ...
Page 130: ...124 Spectrospin AG Daedalus Lock Manual ...
Page 131: ...Daedalus Lock Manual Spectrospin AG 125 Lock Receiver Index D E ...
Page 132: ...126 Spectrospin AG Daedalus Lock Manual ...
Page 133: ...Daedalus Lock Manual Spectrospin AG 127 19F Receiver Option Index A ...
Page 134: ...128 Spectrospin AG Daedalus Lock Manual ...
Page 135: ...Daedalus Lock Manual Spectrospin AG 129 Lock Controller Index A ...
Page 136: ...130 Spectrospin AG Daedalus Lock Manual ...
Page 137: ...Daedalus Lock Manual Spectrospin AG 131 Lock Controller Index B C ...
Page 138: ...132 Spectrospin AG Daedalus Lock Manual ...
Page 139: ...Daedalus Lock Manual Spectrospin AG 133 Lock RS232 Biggy Board ...
Page 140: ...134 Spectrospin AG Daedalus Lock Manual ...
Page 141: ...Daedalus Lock Manual Spectrospin AG 135 Z0 Compensation Option ...
Page 142: ...136 Spectrospin AG Daedalus Lock Manual ...
Page 143: ...Daedalus Lock Manual Spectrospin AG 137 Cables ...
Page 144: ...138 Spectrospin AG Daedalus Lock Manual ...
Page 145: ...Daedalus Lock Manual Spectrospin AG 139 Test Equipment Index A ...