121
Bruker Optik GmbH
HYPERION User Manual
Options and Accessories 5
5.4.3
Installation location
At the left microscope side, there are three slide-in slots for polarizers:
In slot A in fig. 5.6, the analyzer for transmission and reflection measurements can be
inserted.
For viewing the sample surface in reflection mode with crossed polarizers, a second
polarizer needs to be inserted into slot B in fig. 5.6.
For sample viewing in transmission mode, a second polarizers needs to be inserted into
slot C in fig. 5.6.
For IR spectroscopic measurements, usually one IR polarizer is sufficient. The IR polar-
izer can be inserted into any slot (A, B or C in fig. 5.6).
The polarizer holders are interchangeable. They can be inserted into any slot (A, B or C
in fir. 5.6).
WARNING
Health hazard because of improper handling of broken polarizer mate-
rial
Non-observance of the following safety instructions could result in death or serious
health problems (e.g. poisoning symptoms).
➣
Avoid generating dust of broken polarizer material. This material is toxic if swal-
lowed or inhaled.
➣
Avoid skin and eye contact.
➣
Dispose the toxic material according to the laboratory regulation and national reg-
ulations.
➣
Observe the safety instructions of the attached safety data sheet.
Figure 5.6:
Slide-in slots for polarizers
B
A
C
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...