163
Bruker Optik GmbH
HYPERION User Manual
Specifications A
A.5
Detectors
Detectors for HYPERION 1000 and HYPERION 2000
The detectors listed in the table below are single-element detectors.
Detector
Mode of cooling /
nominal hold time
Spectral range
Detection sensitivity
Standard MCT detector
mid band, D316025
with liquid N
2
8 h
12,000 - 600 cm
-1
D*>2x10**10cm Hz
1/2
W
-1
optionaler MCT detector
mid band, D316025MU
with liquid N
2
> 12 h
12,000 - 600 cm
-1
D*>2x10**10cm Hz
1/2
W
-1
optionaler MCT detector
narrow band, D313025
with liquid N
2
8 h
10,000 - 780 cm
-1
D*>4x10**10cm Hz
1/2
W
-1
optionaler MCT detector
narrow band, D313025M
with liquid N
2
> 12 h
10,000 - 780 cm
-1
D*>4x10**10cm Hz
1/2
W
-1
optionaler MCT detector
broad band, D315025
with liquid N
2
8 h
10,000 - 450 cm
-1
D*>5x10**9cm Hz
1/2
W
-1
optionaler MCT detector
broad band, D315025M
with liquid N
2
> 12 h
10,000 - 450 cm
-1
D*>5x10**9cm Hz
1/2
W
-1
InSb-detector (optional)
near-infrared region,
D413
with liquid N
2
8 h
10,000 - 1,850 cm
-1
D*: >1.5x10**11cm Hz/W
InGaAs-detector
(optional), D424/3
No cooling required.
Detector operates at
ambient temperature.
12,800 - 5,800 cm
-1
NEP: <2*10**-14W/Hz½
Si diode (optional)
visible region,
D510/3
No cooling required.
Detector operates at
ambient temperature.
25,000 - 9,500 cm
-1
NEP: <1*10**-14W/Hz½
DLaTGS-detector
(optional), D301/B
Note:
This detector can
be installed in place of
the standard MCT detec-
tor. This detector allows
for measurement with a
spatial resolution of only
> 30 µm.
No cooling required.
Detector operates at
ambient temperature.
12,000 - 350
cm
-1
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...