HYPERION User Manual
Bruker Optik GmbH
Overview 3
3.6
Condenser
☞
For detailed information about how to adjust the condenser, see section 4.6.
Figure 3.9:
Condenser (upper image: left side view, lower image: right side view)
D
A
C
D’
C’
B
Fig. 3.9
Component and Explanation
A
Pinhole aperture for adjusting the condenser
Note: The condenser is adjusted exclusively in the transmission mode.
B
Condenser
C and C’
Condenser focussing knob:
to raise or lower the condenser in relation
to the sample or sample stage in order to maximize the beam intensity in
transmittance mode.
D and D’
Centering screws:
to center the condenser by moving it in x- and/or y-
direction.