52
HYPERION User Manual
Bruker Optik GmbH
Overview 3
15
Visible light source for transmittance mode.
16
Three-position motorized mirror.
Position 1: Only visible light can pass. (hole)
Position 2: Only IR light can pass. (mirror)
Position 3: Both visible light and IR light can pass (Option A689-R). This option
includes a beamsplitter.
17
Pinhole aperture for adjusting the condenser (in transmission mode only)
18
Iris aperture for setting the Koehler illumination in transmittance mode: to enhance
the image contrast
19
Condenser (adjustable in height): to maximize the light intensity on the sample in
transmittance mode.
20
IR beam coming from the spectrometer.
21
Three-position motorized mirror.
Position 1: Mirror deflects the IR beam to the transmittance optics.
Position 2: Mirror deflects the IR beam to the reflectance optics.
Position 3: No mirror intercepts the IR beam. The beam can go to the external acces-
sory port.
22
Video camera port
23
Optional polarizer (for reflectance mode): either for visible light only or for visible light
and IR light (two polarizers). An analyzer (component no. 25) is required in addition.
(It is part of the VIS polarizer.)
24
Optional polarizer (for transmittance mode): either for visible light only or for visible
and IR light (two polarizers). The IR polarizer (A675-P) requires the holder (A 675)
including the polarizer for visible light.
25
Optional analyzer: either for visible light only or for visible and IR light (two polarizers).
Fig. 3.19
and 3.20
Component and Description
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...