69
Bruker Optik GmbH
HYPERION User Manual
Operation 4
8
Define a measurement area of 100 x 100 µm using an aperture.
➣
The precious procedure depends on whether you work with a validated or a non-
validated analysis system and wether the HYPERION microscope is equipped with a
manual knife-edge aperture (standard) or a motorized knife-edge aperture (option).
Non-validated analysis system and
manual knife-edge aperture:
Define the measurement area as shown
in the lower image using the manual
knife-edge aperture
⑥
.
☞
For information about how to operate
the manual knife-edge aperture, see
section 3.7.3.
Note: 100 microns correspond to 5 scale
lines of cross hairs in the binocular (but
not in the OPUS video view!).
Validated analysis system and pinhole
aperture for reflection mode:
Define the measurement area as shown
in the lower image using the pinhole
aperture for the reflection mode
.
To do
this, swing the aperture in the beam path
by rotating the thumb wheel
⑦
in the
arrow direction of the small circle up to
the stop.
Note: Only validated analysis systems
are equipped by default with this aper-
ture.
The result of this aperture setting is a cir-
cular image of the mirror surface and the
circular measurement area has a diame-
ter of 100 µm.
⑥
⑦
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...