Operation 4
4.8
Checking the detector saturation
General information
Especially when you measure in transmittance or when you measure highly reflecting
samples in reflectance, it is advisable to check the detector saturation before each mea-
surement. If the detector is oversaturated, the IR spectroscopic measurement may not
yield meaningful results.
Procedure
1.
Prepare the microscope for a background or sample measurement in transmittance
or reflectance.
☞
For measurements in transmittance, see section 4.10. For measurements in reflec-
tance, see section 4.11.
2.
Activate the measuring mode by actuating the IR button (C in fig. 3.4) on the operat-
ing panel of the microscope.
3.
Start the OPUS software. Select in the
Measure
menu the
Advanced Measurement
function. Load an adequate experiment file (’*’.xpm).
☞
For information about the available standard experiment files, see appendix B.
4.
Click in the
Measurement
dialog on the
Check Signal
tab. Make sure that the
Spec-
trum
option button is activated. (See fig. 4.1.)
The spectrum in fig. 4.1 results from a normal detector saturation.
Figure 4.1:
Spectrum acquired with a normally saturated detector
Enlarged section
This option but-
ton needs to be
activated.
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...