88
HYPERION User Manual
Bruker Optik GmbH
Operation 4
4.12.4
Setting the viewing mode / measuring mode at the ATR objective
The 20x ATR objective allows for both
sample viewing by means of the micro-
scope and spectroscopic analysis
using the measurement techniques
ATR (attentuated total reflection) or
reflection.
For this purpose, the ATR crystal is
brought either in the upper position
(viewing mode) or in the lower position
(measuring mode).
Setting the viewing mode:
Actuate the rocker switch
①
and push
the outer casing
②
of the ATR objec-
tive all the way up at the same time.
Then, release the rocker switch again.
Setting the measuring mode:
Actuate the rocker switch
①
and pull
the outer casing
②
of the ATR objec-
tive all the way down at the same time.
Then, release the rocker switch again.
Attention:
When setting the measur-
ing mode, take care that the ATR crys-
tal tip
③
does not hit the stage.
Potential risk of damaging the ATR
crystal!
Viewing mode
ATR crystal in
upper position
Measuring mode
ATR-crystal in
lower position
①
②
③
Summary of Contents for HYPERION
Page 1: ...HYPERION User Manual I 24319 ...
Page 56: ...54 HYPERION User Manual Bruker Optik GmbH Overview 3 ...
Page 148: ...146 HYPERION User Manual Bruker Optik GmbH Repair and Maintenance 6 ...
Page 168: ...166 HYPERION User Manual Bruker Optik GmbH Specifications A ...
Page 172: ...170 HYPERION User Manual Bruker Optik GmbH Measurement parameters B ...
Page 174: ...172 HYPERION User Manual Bruker Optik GmbH Spare parts and consumables C ...
Page 175: ...173 Bruker Optik GmbH HYPERION User Manual D System diagram ...
Page 176: ...174 HYPERION User Manual Bruker Optik GmbH System diagram D ...