L-TRX/L-19F
118 / 234
H172203_1_001
11.5.3
L-TRX Specific Error Messages
Error
Description/Measures
L-TRX Power Supply error occurred
External or internal power supply failure.
Check power supply board status.
Check power supply diagnostic.
If the power supply input voltages are within
their limits a hardware failure has occurred.
Replace the L-TRX and/or contact a Bruker
service representative.
L-TRX 160MHz clock missing or L-TRX 320
MHz clock missing
See
.
L-TRX Amplifier: overcurrent error occurred See
L-TRX Amplifier: bias current regulator
underflow error occurred
A hardware failure has occurred.
Contact your local Bruker service
representative.
L-TRX Error: Signal BLKTR-2H was
activated without selecting SEL_2H/DEC
Check wiring (if external power amplifier is
used the BLNKTR-2H~ input must be left
open).
Check experiment setup (pulse program).
Check Lock RTP Configuration.
See
L-TRX DSP signal chain: overflow error
occurred
Contact your local Bruker service
representative.
L-TRX FPGA DCM (PLL) lock error
occurred
Reference clock synchronization failure.
Restart BSMS (power off/on).
Check reference clock (REF_IN J2) from
AV4 Reference board.
If the error remains a hardware failure has
occurred. Contact your local Bruker service
representative.
L-TRX ADC overflow occurred
Reduce Lock Power.
L-TRX Diagnostic ADC busy error occurred
Contact your local Bruker service
representative.
Table 11.4: L-TRX Error Messages
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...