Radiation Shield Temperature Monitoring (MAG-RS)
H172203_1_001
201 / 234
18 Radiation Shield Temperature
Monitoring (MAG-RS)
18.1
Introduction
For RS (refrigerated radiation shield) type of magnet systems the temperature of the
refrigerated radiation shield inside the cryostat must be monitored during operation.
For that two PT 100 temperature sensors are mounted inside the cryostat. The temperature
measurement is realized using a digital adapter that can be connected to the BSMS like other
accessory (e.g. Nitrogen Level Sensor or BSMS/2 VTA).
In case of a failure the operator will receive an alarm by MICS.
18.2
Configurations and Installation
Bruker Part
Number
Name
Marking
Typical usage
[Power or Auxiliary]
a
Z122502
BSMS/2 VTA MAG-RS
MAG-RS
POWER
a
POWER means that this application needs heater power and therefore must be
connected through a cable that contains heater power wires to an output of the
VARIABLE POWER SUPPLY BOARD (VPSB(-DC))
Table 18.1: Available Unit
Necessary cable set: Z119854 CABLE SET BSVT AUXILIARY HEATER
18.2.1
Connection to the Console (BSMS/2 with BSVT)
Figure 18.1: Connection to BSMS (Monitoring and Service)
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...