Radiation Shield Temperature Monitoring (MAG-RS)
H172203_1_001
203 / 234
18.2.3
Measurements Provided for Diagnostics
The on-board diagnostics supervise essential board functions like power supply and ADC
state. In case of a failure the board will reset and reboot.
The software running on the ELCB may notify the user about abnormal events.
Status/Errors
The MAG-RS unit can perform the following checks:
• Communication and functionality of the ADC’s.
• Power voltages.
• Shorted or disconnected lines at sensor interface connectors.
18.2.4
Calibration
Factory calibration is stored on the board (no field calibration necessary).
18.2.5
Connectors and LED‘s
See
.
18.2.6
Service
This unit does not need any calibration in the field, the product is factory calibrated.
18.2.6.1
Service Web
For a pulse tube replacement or service, a heater must be activated.
There is an ELCB Service Web page available for service. Please contact your local Bruker
office.
18.2.6.2
Diagnostics and Troubleshooting
The device state is displayed using the 5 LED. See
LED‘s CH1/2/3/4 and VT POWER [
for a detailed description.
In addition, on-board diagnosis data or failure events are sent to the ELCB immediately and
displayed within TopSpin GUI or logfile.
18.2.7
System Requirements
See
18.2.8
Ordering Information
Please contact your local Bruker office.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...