Technical Data
H172203_1_001
219 / 234
25 Technical Data
25.1
General Information
Data
Value
Unit
Weight without subunits
5
kg
Weight with subunits
40
kg
Length
51,0
cm
Width
48,3
cm
Height
31,0
cm
Sound Pressure Level
a
, maximum
63
dBA
a
Measured stand-alone, outside of the cabinet. For SPL including the cabinet and
protection instructions see the applicable manuals.
Table 25.1: Technical Data: General Information
25.2
Connection Values
The BSMS System for AVANCE NEO has no single power inlet with mains distribution within
the chassis. Each power supply module with mains input is connected individually to the PDU
(power distribution unit) of the cabinet. The connecting values are given per supply module.
Data
Value
Unit
Voltage range
a
208 - 230
Vac
Apparent power consumption, rated maximum
8
A
Circuit protection
b
10
Amps AC
Frequency
50/60
Hz
a
Not all units within the cabinet allow input voltages as low as 100 VAC.
Therefore, the usable PSM input range is smaller than on the rating plate of the
AC/DC supply.
b
Two not replaceable internal fuses (L+N) per AC/DC supply, two supplies per
PSM.
Table 25.2: Electrical Connection Values: AV4 PSM-48V
Data
Value
Unit
Voltage
100 - 240
Vac
Apparent power consumption, rated maximum
4
A
Circuit protection
a
4
Amps AC
Frequency
47 - 63
Hz
a
Two replaceable fuses (L+N), time-lag T.
Table 25.3: Electrical Connection Values: AV4 PSB-12V
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...