AV4 BSMS System
30 / 234
H172203_1_001
4.1.2
Software Control of AV4 BSMS
All subunits may be accessed by a Web-based service tool, making service handling easy
and comprehensive.
The ELCB has an automatic save configuration mechanism and stores all the parameters (e.
g. for Lock, Shim, Lift, HE-Level measurement) within its nonvolatile memory (NVM). To be
able to switch back to a known state, a saved configuration from the installation may be
restored later - this fail-safe configuration can be re-activated by any user.
All activities of the AV4 BSMS and the data exchange with the TopSpin application are
logged by the ELCB software. This information is accessible by the service tool and,
additionally, it is periodically transferred to the workstation to keep a detailed long-term
history for trouble shooting. It is configurable how detailed the activities are logged.
4.2
Rack Configurations
In the following sub chapters, there are the various standard configurations described in
detail. For a list of the part numbers of the electronics units, please refer to
.
NOTICE
Units of previous BSMS generations that are not explicitly mentioned in this
manual are not compatible to the AV4 BSMS Chassis!
u
VME based Devices and the old Lock RF Units L-TX and L-RX will be destroyed when
powered up in the AV4 BSMS Chassis.
u
Examples of VME Devices: SLCB, SLCB/2, SLCB/3, PNK, BSMS 2HTX, GAB, SCB13.
AV4 Units shall only be used in the AVANCE NEO generation of NMR Spectrometers.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...