ELCB
54 / 234
H172203_1_001
Parameter
Min
Factory
Default
Max
Unit
Notes
Current Noise
400
nA (pp)
4)
Current Source
Bandwidth
600
Hz
5)
Lock Hold active
100
µs
Lock Hold inactive
100
µs
Lock Hold latency
100
µs
Table 8.1: Parameter and Technical Data
General Note: Any of the values listed above may change without notice.
1. Values from -1000.0 to +1000.0 degrees are accepted, however the actual phase is
evaluated modulo 360 degrees.
2. The actual range depends on the hardware code and on the frequency: The first series of
RF boards (HW code 0) had a range from -50 to +10 dB, the next following series (HW
code 1 and higher) had a range from -60 to +0 dB. For frequencies of 600 MHz and
above, this range has been extended (-60 dB to +10 dB) on the hardware versions 6 and
higher.
3. This value is used for static drift compensation with manual evaluation/definition of the drift
rate (called
Manual Drift Compensation
, e. g. in solids configuration). Static/manual drift
compensation is active while the lock is not sweeping.
4. Maximum Noise between 0.01 Hz and 10 Hz
5. Minimum range of -3 dB point
8.3
Configuration and Wiring
The ELCB is plugged into Slot 10, see
. The drawing below shows the front
panel of an ELCB with the LED‘s and connectors. For proper operation with TopSpin, at least
the Ethernet, the 10 MHz reference clock and the RCP input have to be connected
accordingly.
There is a dedicated connector for the communication with an optional SampleCase
automated sample changer.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...