ELCB
H172203_1_001
61 / 234
8.6
Service Software
For service purpose, there is a Web access available (setup, calibration and diagnostic).
Some of these Web functions are open for all users (e. g. clients), other functions are
reserved for service engineers - it is necessary to log in and enter the required password
before these functions can be accessed (description in the BSMS Service Web chapter).
8.6.1
Lock Service Web
The Submenu
Main
|
Lock
provides access to all service functions in connection with the
ELCB and the related RF boards (L-TRX, L-19F).
Figure 8.6: Main Window for NMR Lock Functions
Most of the functions under the menu point
Lock Control
are normally handled by the
TopSpin application over the CORBA interface. It is however possible to invoke all of these
functions by the Service Web. Also the solvent specific parameters that are normally passed
by the TopSpin application (e. g. Lock Power, Lock Phase, Lock Gain, Loop Gain, Loop
Time, Loop Filter, ..) can be checked and / or defined there.
The point
Lock Configuration
provides setup of the Lock relevant parameters at installation.
Statistics
displays information about Lock failures (in case that the Lock got lost), and
Diagnostics
provides the sub-menu displaying of the
2
H spectrum captured during the last
Auto Lock trial, a sub-menu for configuration and troubleshooting of the RF boards and a
sub-menu containing the history of the Lock Regulator/Drift Compensator.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...