ELCB
68 / 234
H172203_1_001
8.6.4
Part Numbers
Bruker Part No.
Description
Z100818
BSMS/2 ELCB EXTENDED LOCK CTRL BOARD
Table 8.2: Part Numbers: ELCB
8.6.5
Diagnostic and Troubleshooting
If there is a problem related to the Lock, the logging can be configured in the Service Menu
for issuing detailed information about the Lock System. It must be made sure, that the
2
H
path is correctly initialized by typing
ii
in the TopSpin. Additionally, all RF board tests can be
run on the Lock
L-TRX Diagnostic Functions
menu.
Figure 8.12: Lock Diagnostics Menu
Since the Lock needs almost the complete spectrometer for correct operation, it is sometimes
difficult to find a Lock related error. If it cannot lock in or if there are no
Lock wiggles
available
on the Lock display, there are many possible reasons - the field could be completely out of
range, the shims could be erroneously set, there could be a problem on the
2
H path/probe,
some Cryoshims or even the magnet could have quenched.
For checking the magnet (if the
2
H FFA spectrum after a failed Auto Lock trial is completely
flat), it is recommended to run a simple 1H experiment on a water sample with a very large
bandwidth - if there is a peak at all, it will show up.
The diagram below shows the L-TRX Diagnostic page, indicating self-test results and several
voltage and temperature measurements.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...