SCB20
78 / 234
H172203_1_001
2
Some BOSS files provide more than one mode, which can be selected here.
However, it is no longer necessary to differentiate between US and non-US
systems, install mode and user mode. Typically, one mode is sufficient.
3
Here is the identification of the connected Shim System, which is based on 2
(BOSS1) or 3 hardware codes.
4
The maximum power dissipated in the Shim System - this value is defined in the
BOSS file.
5
For specific situations it is possible to adapt/extend the
Shim Power Limit
.
Setting this flag to
yes
allows to override the value of the BOSS file by a user
specific limit (service account necessary) -> the value can then be entered under
point 4.
6
Time in seconds for softly starting up / shutting down the Shim Subsystem.
7
Normally the TopSpin application makes sure that there is an appropriate BOSS
file loaded - it can be however necessary to do this manually. This menu point
provides selecting a specific BOSS file and loading it to the BSMS. The BOSS file
is handled by the ELCB firmware - if the file is not valid (e. g. syntax errors,
missing definitions), the Logging can be checked for details (kind of error, line
number of the BOSS file where the error has been detected).
9.5.3
View and Modify the Shims
Figure 9.6: View and Manual Modification of the Shims
The example in the diagram above shows a Shim Subsystem configured for 28 available/
accessible Shims. By the small panel at the bottom, the selected Shim (
Z
in our example) can
be viewed or modified manually.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...