6.0 Working with the stereo microscope 
 

  Set the focusing (F) on neutral position. 

  Place an object onto the working space and adjust the height of the horizontal arm (J) until 

you have a passably sharp image , than fix the height of horizontal arm (J) by the locking 
knob (N). 

  Set  the  diopter  correction  (B)  of  both  eyepieces  in  the  neutral  position  (marking  notch  on 

tube) 

  Adjust  the  interpupillary  distance  by  moving  both  prism  housings  (C)  towards  each  other 

until one round image is seen with both eyes. 

 
 
6.1 Setting the eyepieces in order to have a sharp image over the complete zoom range 

 

  Set  the  magnification  at  the  highest  position  by  means  of  the  zoom  knob  (D)  and  sharply 

focus on a flat specimen. 

  Now bring the magnification down to the lowest position. 

  First  correct  the  sharpness  with  only  the  left  hand  diopter  correction  (B),  and  than  for  the 

right hand one.  

 
 
 
7.0 Maintenance and cleaning 

 
Always place the dustcover over the microscope after use. Keep the eyepieces always mounted in 
the tubes to avoid dust entering the instrument.

 

 
 
7.1 Cleaning the optics  

 
When the eyepiece lens is dirty it can be cleaned by wiping a piece of lens paper over the surface 
(circular movements). When this does not help put a drop of xylol or alcohol on the lens paper. 
Never put xylol or alcohol directly on the lens! 
 
When dirt is clearly visible in the field of view it resides on the lowest lens of the eyepiece. By 
using the Allen-key the eyepiece can be removed from the tube. Clean the outside of the lens. 
In case there is still dust visible please check if the dust is in the eyepiece by turning it. If this is the 
case remove the lowest lens carefully from the eyepiece and clean it.  

 
Caution  

 

  Cleaning cloths containing plastic fibres can damage the coating of the lenses! 

 
 