4. MODIFYING THE MEASUREMENT CONDITIONS
●
Operating procedure
[MODE/ESC]
2
[n/ENT]
[n/ENT]
3
1
4
A
B
4
5
[CUTOFF
↑
] or [
↓
]
[CUTOFF
↑
] , [
↓
] or
[REMOTE
←
]
4.2
4.3
[n/ENT]
4.1
[n/ENT]
[n/ENT]
[PARAMETER]
Count level for
Pc or Sm not set
Count level for
Pc or Sm set
μ
%
%
%
%
Ra
m
μ
m
μ
m
In the measurement mode press the [MODE/ESC] key.
This switches to the parameter calculating condition setup
mode.
If the count level for Pc or Sm is not set, follow sequence A
shown at the left.
<Sequence A>
Press the [PARAMETER] key.
If the count level for Pc or Sm is set, follow sequence B
shown at the left.
<Sequence B>
Press the [n/ENT] key.
Specify the unit of the count level for Pc or Sm.
Each time the [CUTOFF
↑
] or [
↓
] key is pressed the unit of
the level toggles between “%” and “
μ
m (
μ
in)”.
Here, the explanation assumes that “%” is used as the unit of
the level.
Enter the count level.
The permissible range is as follows:
1 to 99% (at 1% increments)
0.1 to 100.0
μ
m (at 0.1
μ
m increments)
or 5 ~ 4000
μ
in (at 1
μ
in increments)
TIP
For information on how to enter a numeric
value refer to “2.2 Key Operation to Enter
Numeric Values”.
After the desired count level is specified, press the [n/ENT] key.
The setting of the count level is identical for both Pc and Sm,
since they are reciprocals of one another.
Therefore, if the Sm is selected, “Pc” is displayed in the
setting display.
r
e
t
e
m
a
r
a
P
y
a
l
p
s
i
d
g
n
i
t
t
e
S
c
P
r
m
c
P
䊊
-
m
S
䊊
-
r
m
-
䊊
Summary of Contents for 148 460
Page 8: ...No 99MBB079A MEMO vii ...
Page 26: ...1 18 No 99MBB079A MEMO ...
Page 134: ...6 36 No 99MBB079A MEMO ...
Page 140: ...7 6 No 99MBB079A MEMO ...
Page 144: ...8 4 No 99MBB079A MEMO ...