INTERNATIONAL ELECTRICAL SYMBOLS 

 

!

 

Caution !  Refer to the explanation in this Manual 

 

 

Caution !  Risk of electric shock 

 

 

Earth (Ground) 

 

 

Double Insulation or Reinforced insulation 

 

 

Fuse 

 

 

AC--Alternating Current 

 

 

DC--Direct Current 

 

3) CENELEC DIRECTIVES 

The  instruments  conform  to  CENELEC  Low-voltage  directive  2006/95/EC  and  Electromagnetic  compatibility 
directive 2004/108/EC 

 

4)  OPERATION  

Note: All function operations described hereafter are via the Red test lead for positive (+) polarity and Black test 
lead for Ground reference (-), or otherwise specified 

Warning: Before and after hazardous voltage measurements, test the voltage function on a known source such as 
line voltage to determine proper meter functioning. 

 

4-

1) FUNCTIONS IN “Auto V.

” POSITION (

Models 27s, 25s 

& 25a

 

only) 

-Quick Start (

Models 27s

25s 

& 25a

 only) 

AutoCheck

TM

 mode is the default function in

 

Auto V.

” position. 

Press the SELECT button momentarily to select 

and step through the functions: 

AutoCheck

TM

→Continuity→EF→ACV→DCV→

→Hz→Cx→AutoCheck

TM

 

 

-AutoCheck

TM

 mode (

Model 27s, 25s 

& 25a

 only) 

This innovative AutoCheck

TM

 feature automatically selects measurement function of DCV, ACV or Resistance (

based on the input via the test leads. 

●With no input, the meter displays “Auto” when it is ready.

 

With no voltage signal but a resistance below 6M

 is present, the meter displays the resistance value. When 

below 25

 is present, the meter (

Model 27s

 only) further gives a continuity beep tone. 

When a signal above the threshold of DC 1.2V or AC 1.5V up to the rated 600V (450V for 

Model 25s

 & 25a

) is 

present, the meter displays the appropriate voltage value in DC or AC, whichever larger in peak magnitude. 

Overload-Alert  Feature:

 When above rated 600V (450V for 

Model 25s

 & 25a

) is present, the meter displays 

“OL” with a warning beep tone. Disconnect the test leads from the signal immediately to avoid h

azards. 

Note: 

*Range-Lock  Feature:

  When  a  measurement  reading  is  being  displayed  in  AutoCheck

TM

  mode,  press  the 

SELECT

 button momentarily 

1 time

 can lock the function-range it was in. The LCD annunciator 

“ 

 turns 

off. Range-lock can speed up repetitive measurements. Press the 

SELECT

 button momentarily again to return to 

AutoCheck

TM

 mode.

 

*Hazardous-Alert:

  When  making  resistance  measurements  in  AutoCheck

TM

  mode,  an  unexpected  display  of 

voltage readings alerts you that the part under test is still energized. 

*

AutoCheck

TM

 nominal input impedances are slightly lower than that of common digital multimeters. They are, in 

fact, more similar to that of most traditional analog multimeters. “LoZ” will display on the LCD to remind the users 

in such cases. Although not likely, if such impedance might damage your circuits, use the common impedance (Hi-
Z) voltage mode (

Models 27s & 22s

 only) for making voltage measurements on them. 

 

-Continuity, Audible With Symbolic Display (

Models 27s, 25s 

& 25a

 

only) 

From AutoCheck

TM

 

with “Auto” being displayed, p

ress the 

SELECT

 button momentarily 

1 time

 to select Continuity 

function. The meter displays a symbolic open-switch display 

” when it is ready.

 Continuity is convenient for 

checking  wiring  connections  and  operation  of  switches.  A  continuous  beep  tone  with  a  symbolic  closed-switch 
display 

” indicates a complete circuit. In noisy environments, it is helpful to “see” continuity measurements.

 

Summary of Contents for BM22s

Page 1: ... 1 Ed 3 0 IEC EN61010 2 030 Ed 1 0 IEC EN61010 2 033 Ed 1 0 IEC UL EN61010 031 Ed 1 1 and CAN CSA C22 2 No 61010 1 12 Ed 3 0 to CAT III 300V CAT II 600V CAT II 450V for Model 25s 25a In case of removing test lead tip insulation CAP to CAT II 600V CAT II 450V for Model 25s 25a only Measurement Category III CAT III is for measurements performed in the building installation Examples are measurements ...

Page 2: ...25s 25a is present the meter displays the appropriate voltage value in DC or AC whichever larger in peak magnitude Overload Alert Feature When above rated 600V 450V for Model 25s 25a is present the meter displays OL with a warning beep tone Disconnect the test leads from the signal immediately to avoid hazards Note Range Lock Feature When a measurement reading is being displayed in AutoCheckTM mod...

Page 3: ...s DCV is the default function Press SELECT button momentarily to select ACV The AC annunciator turns on Press momentarily again to activate the Line Level Hz function Note Line Level Hz input sensitivity varies automatically with ACV range selected when Line Level Hz is selected AC 6V range has the highest and AC 600V range has the lowest sensitivity Measuring the signal in ACV function WHILE sele...

Page 4: ...ily AGAIN selects Capacitance function Frequency function Model 22s only Rotate the rotary selector to the Hz position selects Frequency function Unlike the Line Level Hz Frequency function for Model 27s only as stated this Common Hz Frequency function is set only at the highest input sensitivity mainly for measuring small electronic signals of below 20VAC rms 4 3 OTHER FEATURES Power On Off For M...

Page 5: ...ect test leads from live circuits Loosen the screw from the case bottom Lift the end of the case bottom nearest the input test leads until it unsnaps from the case top Replace the battery Observe battery polarities with positive faces up towards the case bottom Replace the case bottom and ensure that the snap on the case top near the LCD side is engaged Re fasten the screw Note for Models 27s 25s ...

Page 6: ...del 22s Data Hold and 30ms MAX Hold Model 25a 25s 27s AutoCheckTM Automatic V Selection and EF Detection Accessories Battery installed and User s manual Optional Accessories BH 20R protective holster and BSC 20 soft carrying pouch Electrical Specification Accuracy is given as of reading digits number of digits or otherwise specified 23o C 5o C and less than 75 R H DC Voltage Models 25a 25s 27s RAN...

Page 7: ...nge accuracy is not specified 2 Accuracies with film capacitor or better 3 Updates 1 minute on large values 4 Specified with battery voltage above 2 8V half full battery Accuracy decreases gradually to 12 at low battery warning voltage of approx 2 4V Ohms Models 25a 25s 27s RANGE1 Accuracy 600 0 2 0 6d2 6 000k 1 2 6d2 60 00K 600 0K 1 0 4d 6 000M 2 0 4d Open Circuit Voltage 0 4VDC typical 1 AutoChe...

Page 8: ...s warranty does not apply to accessories fuses fusible resistors spark gaps batteries or any product which in BRYMEN s opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation or handling To obtain warranty service contact your nearest BRYMEN authorized agent or send the product with proof of purchase and description of the difficulty postage and insura...