background image

 

Measurement Category:  

Models 22s & 27s:

 CAT II 600V and CAT III 300V  

Model 25s

 & 25a

: CAT II 450V and CAT III 300V 

E.M.C.:

 Meets EN61326 (1997, 1998/A1), EN61000- 4-2 (1995), & EN61000-4-3 (1996) 

In an RF Field of 3V/m: 

Capacitance function is not specified 
Other function ranges:  
Total accuracy = Specified ac 45d 

Performance above 3V/m is not specified 

Overload Protection: 

Models 22s & 27s

920VDC & VAC rms

  

Model 25s 

& 25a

855VDC & VAC rms

  

Low Battery:

 Below approx. 2.4V 

Power Supply:

 3V standard button battery x 1 (IEC-CR2032; ANSI-NEDA-5004LC) 

Power Consumption (typical):  

Model 22s, 25s

 & 25a

: 2mA 

Model 27s

: 6mA for Voltage functions on Auto-V

 position, and 2mA for other functions 

APO Consumption (typical):

 2.2

A  

APO Timing:  

Model 22s

:  Idle for 30 minutes 

Model 

25a, 

25s & 27s

: Idle for 3 minutes 

Dimension / Weight:

 L113mm x W53mm x H10.2mm / Approx. 78 gm 

Special Features: 

Model 22s

: Data Hold, and 30ms MAX Hold 

Model 

25a,

 25s & 27s

: AutoCheck

TM

 (Automatic V & 

 Selection), and EF-Detection 

Accessories:

 

Battery installed, and User’s manual

 

Optional Accessories:

 BH-20R protective holster, and BSC-20 soft carrying pouch 

 

Electrical Specification 

Accuracy is given as +/- (% of reading  number of digits) or otherwise specified @ 23

o

C +/- 5

o

C and less 

than 75% R.H. 

-----------------------------------------------------------------------      ------------------------------------------------------------------------ 
DC Voltage (

Models 

25a, 

25s & 27s

RANGE 

Accuracy 

6.000V 

0.5%+3d

 

60.00V 

1.0%+5d 

450.0V

 1)

 

1.2%+5d 

600.0V

 2)

 

2.0%+5d 

600.0V

 3)

 

1.5%+5d 

BM25 Input Impedance: 

AutoCheck

TM

 Lo-Z DCV: 160k

, 160pF nominal 

BM27 Input Impedance: 

AutoCheck

TM

 Lo-Z DCV: 833k

 (4.2k

 when 

displaying “Auto”), 90pF nominal

 

Hi-Z DCV: 5M

, 90pF nominal 

NMRR: > 30dB @ 50Hz/60Hz 
CMRR: > 100dB @ DC, 50Hz/60Hz; Rs=1k

  

 

DCV AutoCheck

TM

 Threshold: 

> +1.2VDC or < -0.6VDC nominal 

1)

BM25 AutoCheck

TM

 DCV only 

2)

BM27 AutoCheck

TM

 DCV only 

3)

BM27 Hi-Z DCV only 

 

DC Voltage (

Model 22s

)  

RANGE 

Accuracy 

400.0mV 

1.0%+2d

 

4.000V, 40.00V, 400.0V 

2.0%+2d 

600V 

2.5%+4d 

NMRR: > 50dB @ 50Hz/60Hz  
CMRR: > 120dB @ DC, 50Hz/60Hz; Rs=1k

  

Input Impedance: 10 M

, 30pF nominal;  (1000M

 

for 400.0mV range) 

Summary of Contents for BM22s

Page 1: ... 1 Ed 3 0 IEC EN61010 2 030 Ed 1 0 IEC EN61010 2 033 Ed 1 0 IEC UL EN61010 031 Ed 1 1 and CAN CSA C22 2 No 61010 1 12 Ed 3 0 to CAT III 300V CAT II 600V CAT II 450V for Model 25s 25a In case of removing test lead tip insulation CAP to CAT II 600V CAT II 450V for Model 25s 25a only Measurement Category III CAT III is for measurements performed in the building installation Examples are measurements ...

Page 2: ...25s 25a is present the meter displays the appropriate voltage value in DC or AC whichever larger in peak magnitude Overload Alert Feature When above rated 600V 450V for Model 25s 25a is present the meter displays OL with a warning beep tone Disconnect the test leads from the signal immediately to avoid hazards Note Range Lock Feature When a measurement reading is being displayed in AutoCheckTM mod...

Page 3: ...s DCV is the default function Press SELECT button momentarily to select ACV The AC annunciator turns on Press momentarily again to activate the Line Level Hz function Note Line Level Hz input sensitivity varies automatically with ACV range selected when Line Level Hz is selected AC 6V range has the highest and AC 600V range has the lowest sensitivity Measuring the signal in ACV function WHILE sele...

Page 4: ...ily AGAIN selects Capacitance function Frequency function Model 22s only Rotate the rotary selector to the Hz position selects Frequency function Unlike the Line Level Hz Frequency function for Model 27s only as stated this Common Hz Frequency function is set only at the highest input sensitivity mainly for measuring small electronic signals of below 20VAC rms 4 3 OTHER FEATURES Power On Off For M...

Page 5: ...ect test leads from live circuits Loosen the screw from the case bottom Lift the end of the case bottom nearest the input test leads until it unsnaps from the case top Replace the battery Observe battery polarities with positive faces up towards the case bottom Replace the case bottom and ensure that the snap on the case top near the LCD side is engaged Re fasten the screw Note for Models 27s 25s ...

Page 6: ...del 22s Data Hold and 30ms MAX Hold Model 25a 25s 27s AutoCheckTM Automatic V Selection and EF Detection Accessories Battery installed and User s manual Optional Accessories BH 20R protective holster and BSC 20 soft carrying pouch Electrical Specification Accuracy is given as of reading digits number of digits or otherwise specified 23o C 5o C and less than 75 R H DC Voltage Models 25a 25s 27s RAN...

Page 7: ...nge accuracy is not specified 2 Accuracies with film capacitor or better 3 Updates 1 minute on large values 4 Specified with battery voltage above 2 8V half full battery Accuracy decreases gradually to 12 at low battery warning voltage of approx 2 4V Ohms Models 25a 25s 27s RANGE1 Accuracy 600 0 2 0 6d2 6 000k 1 2 6d2 60 00K 600 0K 1 0 4d 6 000M 2 0 4d Open Circuit Voltage 0 4VDC typical 1 AutoChe...

Page 8: ...s warranty does not apply to accessories fuses fusible resistors spark gaps batteries or any product which in BRYMEN s opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation or handling To obtain warranty service contact your nearest BRYMEN authorized agent or send the product with proof of purchase and description of the difficulty postage and insura...

Reviews: