14
GENERAL SPECIFICATION
Display:
3-5/6 digits 6,000 counts & 4-1/2 digits 20,000 counts max
Polarity:
Automatic
3-5/6 digits Update Rate:
5 per second nominal
4-1/2 digits Update Rate:
5 per second nominal
61 Segments Bar graph Update Rate:
40 per second max
Operating Temperature:
-10
C to 50
C
Relative Humidity:
Maximum relative humidity 90% for temperature up to 28
C
decreasing linearly to 50% relative humidity at 50
C
Pollution Degree:
2
IP Rating:
IP40
Storage Temperature:
-20
C to 60
C, < 80% R.H. (with battery removed)
Altitude:
Operating below 2000m
Temperature Coefficient:
nominal 0.1 x (specified accuracy)/
C @(-10
C ~ 18
C or
28
C ~ 50
C), or otherwise specified
Sensing:
AC True RMS
Safety:
ETL certified per IEC/UL/EN61010-1 Ed. 3.0, IEC/UL/EN61010-2-030 Ed. 1.0,
IEC/UL/EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and the corresponding
CAN/CSA-C22.2 regulations to Measurement Categories:
CAT III 1000 V AC & DC and Category IV 600V AC & DC
Overload Protections:
µA & mA: 0.4A/1kV DC/AC rms, IR 30kA, F Fuse; or better
A: 11A/1kV DC/AC rms, IR 20kA, F Fuse; or better
V: 1100V DC/AC rms
mV,
& Others: 1000 V DC/AC rms
Transient Protection:
8kV (1.2/50
s surge)
E.M.C.:
Meets EN61326-1:2013
ACV of Model 836, in an RF field of 1V/m:
Total Accuracy = Specified Ac 30 digits
Performance above 1V/m is not specified
All Other Functions, in an RF field of 3V/m:
Total Accuracy = Specified Ac 150 digits
Performance above 3V/m is not specified
Power Supply:
Four Alkaline AA batteries (IEC LR6)
Power Consumption:
8mA typical for VFD ACV
+Hz
; 6.5mA typical for other functions
Low Battery:
approx. 4.6V
APO Timing:
Idle for 20 minutes