16
2)
Add 20d @ reading >80% of range
3)
Unspecified for readings < 5% of range
4)
Signal peak absolute values, including DC bias, less than 1100mV
peak
5)
40Hz ~ 500Hz only for model 836
AC Voltage (20000-count mode)
RANGE
Accuracy
Model 839
Model 836
50Hz ~ 60Hz
199.99mV
3)
, 1.9999V, 19.999V, 199.99V, 1000.0V
0.7% + 30d
0.5% + 30d
40Hz ~ 1kHz
199.99mV
3)
, 1.9999V, 19.999V, 199.99V
1 % + 30d
1000.0V
4)
2 % + 30d
1kHz ~ 5kHz
199.99mV
3)
, 1.9999V, 19.999V, 199.99V
1.8% + 40d
2% + 40d
1)
1000.0V
Unspecified
5kHz ~ 20kHz
2)
199.99mV
3)
, 1.9999V, 19.999V, 199.99V
2% + 40d
Unspecified
1000.0V
Unspecified
Input impedance: 10M
, 110pF nominal
1)
Add 1% @ bandwidth >4kHz ~ 5kHz
2)
Unspecified for readings < 5% of range
3)
Signal peak absolute values, including DC bias, less than 1100mV
peak
4)
40Hz ~ 500Hz only for model 836
VFD AC Voltage
RANGE
Accuracy
1)
10Hz ~ 45Hz
600.0V, 1000V
4% + 5d
45Hz ~ 200Hz
600.0V, 1000V
2.5% + 5d
200Hz ~ 440Hz
600.0V, 1000V
9% + 5d
2)
Input impedance: 10M
, 110pF nominal
1)
Unspecified for fundamental frequency > 440Hz
2)