20
DC current (20000-count mode)
RANGE
Accuracy
Burden Voltage
Model 839
Model 836
199.99
A
1)
0.18% + 40d
0.36% + 40d
0.2mV/
A
1999.9
A
1)
0.18% + 20d
0. 36% + 20d
0.2mV/
A
19.999mA
1)
0.18% + 40d
0. 36% + 40d
3.0mV/mA
199.99mA
1)
0.18% + 30d
0. 36% + 30d
3.0mV/mA
1.9999A
0.5% + 40d
0.6% + 40d
30mV/A
10.000A
2)
0.7% + 20d
0.8% + 20d
30mV/A
1)
A/mA DC accuracies will be affected by extreme interior temperatures of the meter.
For rated accuracies, allow 6 to 20 minutes cool down interval after measuring A-
currents of 3 to 10A continuously.
2)
10A continuous up to ambient 35
o
C; <15 mins on per >5 mins off @ 35
o
C ~ 50
o
C.
>10A to 20A for <30 seconds on per >5 mins off
AC current (6000-count mode)
RANGE
Accuracy
Burden Voltage
Model 839
Model 836
40Hz ~ 3kHz
600.0
A, 6000
A
1% + 2d
1.2% + 2d
0.2mV/
A
60.00mA, 600.0mA
1)
3mV/mA
6.000A, 10.00A
2)
1% + 4d
1.2% + 4d
30mV/A
3kHz ~ 5kHz
600.0
A, 6000
A
2% + 2d
Unspecified
0.2mV/
A
60.00mA, 600.0mA
1)
3mV/mA
6.000A, 10.00A
2)
Unspecified
30mV/A
1)
400mA continuous; >400mA for <1.1 hours on per >20 minutes off
2)
10A continuous up to ambient 35
o
C; <15 mins on per >5 mins off @ 35
o
C ~ 50
o
C.
>10A to 20A for <30 seconds on per >5 mins off