background image

 

The  instrument  is  protected  throughout  by  Double  Insulation  per  IEC/UL/EN61010-1 
Ed.  3.0,  IEC/EN61010-2-030  Ed.  1.0,  IEC/EN61010-2-033  Ed.  1.0,  IEC/UL/EN61010-
031 Ed. 1.1 and CAN/CSA-C22.2 No. 61010-1-12 Ed. 3.0 to Measurement CAT-III 1kV 
and CAT-IV 600V, AC  &  DC.  All  input  terminals  are  also  rated  to  such  Measurement 
Categories requirements. 
The instrument also meets the relevant Parts of EN61557 for CE requirements, and are 
not certified by UL.. In particular, Part 1 Ed. 2.0 General requirements, Part 2 Ed. 2.0 
Insulation  Resistance  and  Part  4  Ed.  2.0  Resistance  of  earth  connection  and 
equipotential bonding, where applicable. 
 
WARNING  
To  reduce  the  risk  of  fire  or  electric  shock,  do  not  expose  this  product  to  rain  or 
moisture. To avoid electrical shock hazard, observe the proper safety precautions when 
working  with  voltages  above  60  VDC  or  30  VAC  rms.  These  voltage  levels  pose  a 
potential shock hazard to the user. Do not touch test lead tips or the circuit being tested 
while  power  is  applied  to  the  circuit  being  measured.  Keep  your  fingers  behind  the 
finger guards of the test leads during measurement. Inspect test leads, connectors, and 
probes  for  damaged  insulation  or  exposed  metal  before  using  the  instrument.  If  any 
defects  are  found,  replace  them  immediately.  Never  attempt  a  voltage  measurement 
with the test lead inserted into the 

/

mA or 

 input jack that is available. Only 

replace the blown fuse with the proper rating as specified in this manual.

 Only use the 

Test  Probe  Assemblies  provided  with  the  equipment  or  UL  Listed  Probe  Assemblies 
with same rating or better. IEC 61010-031 requires exposed conductive test probe tips 
to  be 

  4mm  for  CAT  III  &  CAT  IV  ratings.  It  is  commonly  achieved  by  permanently 

over-molded  plastic  shrouds,  or  by  detachable  Cap  shrouds  for  interchangeable 
between CAT II ratings. Refer to the category markings  on  your  probe  assemblies  as 
well  as  on  the  add-on  accessories  (like  Caps  or  Clips),  if  any,  for  applicable  rating 
changes. 
 
CAUTION 
Disconnect the test leads from the test points before changing functions. Always set the 
instrument to the highest range and work downward for an unknown value when using 
manual ranging mode. 
 
2) CENELEC DIRECTIVES 
The  instruments  conform  to  CENELEC  Low-voltage  directive  2006/95/EC  and 
Electromagnetic compatibility directive 2004/108/EC

 

Summary of Contents for BM876

Page 1: ...USER S MANUAL BM878 BM877 BM876 Insulation Tester Combination Multimeter ...

Page 2: ... circuit breaker in the building installation Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Examples are measurements on distribution boards including secondary meters circuit breakers wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and equip...

Page 3: ...finger guards of the test leads during measurement Inspect test leads connectors and probes for damaged insulation or exposed metal before using the instrument If any defects are found replace them immediately Never attempt a voltage measurement with the test lead inserted into the mA or input jack that is available Only replace the blown fuse with the proper rating as specified in this manual Onl...

Page 4: ...s identifying potentiometer clicks and indicating signal spikes during adjustments 1 3 5 6 digits 6000 counts dual displays 2 Push buttons for special functions features 3 Selector to turn the Power On or Off and Select a function 4 Input Jack for Earth Continuity Test function 5 Common Ground reference Input Jack for all functions EXCEPT Insulation Resistance function 6 Input Jack for all functio...

Page 5: ...momentarily to toggle the subject functions Last selection will be saved as power up default for repeat measurement convenience For ACV Hz function press the RANGE button momentarily to select other ranges when needed For VFD ACV Hz function only 600V range is available to best cope with the range of most Variable Frequency Drives VFD measurements To select DCV function turn the Rotary Knob to DCV...

Page 6: ...ence Continuity function is convenient for checking wiring connections and operation of switches A continuous beep tone indicates a complete wire CAUTION Using resistance or continuity function in a live circuit will produce false results and may damage the instrument In many cases the suspected component must be disconnected from the circuit to obtain an accurate reading ...

Page 7: ...nt Large value capacitors should be discharged through an appropriate resistance load Normal forward voltage drop forward biased for a good silicon diode is between 0 400V to 0 900V A reading higher than that indicates a leaky diode defective A zero reading indicates a shorted diode defective An OL indicates an open diode defective Reverse the test leads connections reverse biased across the diode...

Page 8: ...7 DCmV ACmV Hz functions Model 878 only Press the SELECT button momentarily to toggle the subject functions Last selection will be saved as power up default for repeat measurement convenience ...

Page 9: ...banana plug type K temperature bead probe Bkp60 with correct polarities You can also use a plug adapter Bkb32 Optional purchase with banana pins to type K socket to adapt other standard type K mini plug temperature probes DCmA ACmA Hz functions Model 878 only Press the SELECT button momentarily to toggle the subject functions Last selection will be saved as power up default for repeat measurement ...

Page 10: ...n the Remote Probe Check the fuse before each If the fuse is open the meter will display OPEn when the is being activated at no circuit connection to the probes Refer to the maintenance section for fuse replacement is inhibited when the meter beeps and displays 2V plus warning against energized circuit of more than 2V is being connected before the is active Connecting to energized circuits when th...

Page 11: ...dicated on the secondary display as 200mA or 90mA The meter further gives a beep sound for continuity when the active Resistance reading is 2Ω Default startup primary display reading is Allow enough time for a good measuring result After the is released the last measuring display stays until the next or a function change Lock Test mode is recommended for continuous measurements To apply press the ...

Page 12: ...eference value To apply activate the Lock Test as mentioned above Connect the probes together to show the residue resistance reading and then press the ZERO button momentarily The LCD will display a zero reading with the annuciator turns on The residue resistance reading is then saved temporarily as offset value for the that follows until a further function change or power off ...

Page 13: ...all only be carried out on parts of an installation or equipment that are de energized Connecting to energized circuits when the is active will produce false results and may damage the instrument Always check with voltage functions and remove power from the circuits before carrying out the Setup as illustrated below Select an intended test voltage of 50V 100V 250V 500V or 1000V For model 878 press...

Page 14: ...xt or a function change The secondary display keeps showing the actual detected voltage readings Lock Test mode is recommended for continuous measurements To apply press the LOCK button momentarily to display the annunciator before pressing the TEST button momentarily The LCD will show both to indicate continuous measurement is active Press again either button momentarily to release the Lock Test ...

Page 15: ...14 ...

Page 16: ...he COMPARE button momentarily The LCD annuciator turns on Press the COMPARE button momentarily again to select the preset threshold values in sequence Last threshold values selected for each test voltage range will be saved separately as power up default for repeat measurement convenience Press the COMPARE button for 1 second or more to exit NOTE Selectable threshold values for the test voltage ra...

Page 17: ...eneral guide to interpret the PI or DAR test results are Insulation Condition PI Polarization Index DAR Dielectric Absorption Ratios Dangerous 1 0 Questionable 2 0 1 3 Good 4 0 1 6 Excellent 4 0 1 6 Press the PI DAR button momentarily to enable PI mode with the LCD annunciator turned on Press momentarily again toggles to the DAR mode with the LCD annunciator turned on The secondary display shows t...

Page 18: ... down The primary display shows the real time resistance readings until the timer shows 00 00 for a result The maximum display result is 5 0 and the display shows 5 0 for results beyond that If the measured resistance reading is over range the meter will stop and display Err Press the PI DAR button for 1 second or more to exit ...

Page 19: ...urns off Press the button momentarily again to select an adjacent range Press and hold the button for 1 second or more to resume auto ranging Hold The hold feature freezes the display for later view Press the HOLD button momentarily to toggle the hold feature This feature does not apply to Earth Continuity Test Insulation Resistance functions MAX MIN Recording mode Press REC button momentarily to ...

Page 20: ...any circuit remove the test leads from the input jacks and turn OFF the meter before opening the case Do not operate with open case Install only the same type of fuse or equivalent Calibration Accuracy is specified for a period of one year after calibration Periodic calibration at intervals of one year is recommended to maintain meter accuracy Cleaning and Storage Periodically wipe the case with a...

Page 21: ...ANTY section for obtaining warranty or repairing service Battery and Fuse replacement Battery use Four 1 5V AA battery IEC LR6 Fuses use Fuse F2 for mA input 0 4A 1000Vac Vdc IR 30kA or better FF fuse Dimension 6 x 32 mm Fuse F3 for Earth Continuity Test input 0 25A 1000Vac Vdc IR 30kA or better FF fuse Dimension 6 x 32 mm Battery and Fuse replacement Loosen the screws from the access cover of the...

Page 22: ... 0 IEC UL EN61010 031 Ed 1 1 and CAN CSA C22 2 No 61010 1 12 Ed 3 0 to Category III 1000 V AC DC and Category IV 600V AC DC Compliance to IEC EN61557 2007 Per CE requirements not certified by UL IEC EN61557 1 IEC EN61557 2 IEC EN61557 4 where applicable Overload Protections Insulation Resistance mA Model 878 only 0 4A 1KV IR 30kA or better Earth Continuity Test Models 878 877 only 0 25A 1KV IR 30k...

Page 23: ...Accessories Test probe pair Alligator clip pair Remote control probe Models 878 877 only Holster User s manual Bkp60 banana plug K type thermocouple Model 878 only Optional Accessories Remote control probe Model 876 only BKB32 banana plug to type K socket plug adaptor Model 878 only Special Features Record MAX MIN readings Display Hold Backlighted LCD VFD V Hz readings Lock Test mode for Insulatio...

Page 24: ...00 0V 1000V Unspecified Input impedance 10M 110pF nominal VFD AC Voltage RANGE Accuracy1 10Hz 45Hz 600 0V 4 5d 45Hz 200Hz 600 0V 2 5d 200Hz 440Hz 600 0V 7 5d2 1 Unspecified for fundamental frequency 440Hz 2 Accuracy linearly decreases from 2 5d 200Hz to 7 5d 440Hz Input impedance 10M 110pF nominal DC Voltage RANGE Accuracy 6 000V 60 00V 600 0V 0 2 3d 1000V 0 3 3d Input impedance 10M 110pF nominal ...

Page 25: ...cal Open Circuit Voltage 2 000V 1 5 4d 0 5mA 2 8 VDC Capacitance Model 878 only RANGE Accuracy1 3 000 F2 30 00 F 300 0 F 3000 F 1 5 5d 30 00mF 10 5d 1 Accuracies with film capacitor or better 2 Readings not available below 200nF DCmV Model 878 only RANGE Accuracy 60 00mV 0 5 3d 600 0mV 0 1 3d Input impedance 10M 140pF nominal ACmV Model 878 only RANGE Accuracy 50Hz 60Hz 1 3d 60Hz 3kHz 2 3d 60 00mV...

Page 26: ...mA 600 0mA 0 5 3d 3 0mV mA ACmA Model 878 only RANGE Accuracy Burden Voltage 50Hz 1KHz 60 00mA 600 0mA 1 5 3d 3 0mV mA Earth Continuity Test Models 878 877 only RANGE Test Current Accuracy Measuring Range1 2 000 200mA 0 015 2 199 20 00 90mA 1 5 3d 0 15 21 99 Open Circuit Voltage 4VDC Live Circuit Detector Inhibit test if terminal voltage 2V prior to initialization of test 1 Specified measuring ran...

Page 27: ...rrent Accuracy 50V 3 000M 30 00M 55 0M 1mA 50k 100V 3 000M 30 00M 110 0M 1mA 100k 250V 3 000M 30 00M 275 0M 1mA 250k 500V 3 000M 30 00M 300 0M 550 0M 1mA 500k 1 5 5d 3 000M 30 00M 300 0M 1 5 5d 3000M 2 0 5d 1000V 25 0G 1mA 1M 10 5d 1 Actual output voltage 100 120 of Test Voltage Live Circuit Detector Inhibit test and display voltage reading instead if terminal voltage 30V prior to initialization o...

Page 28: ...nce prepaid to BRYMEN TECHNOLOGY CORPORATION BRYMEN assumes no risk for damage in transit BRYMEN will at its option repair or replace the defective product free of charge However if BRYMEN determines that the failure was caused by misused altered neglected or damaged by accident or abnormal conditions of operation or handling you will be billed for the repair THIS WARRANTY IS EXCLUSIVE AND IS IN L...

Reviews: