12
Accessories : Test lead pair, batteries installed, user's manual, soft carrying pouch, &
BKP60 banana plug type-K thermocouple (model 063 & 065 only)
Optional accessories : BKB32 banana plug to type-K socket plug adaptor (model 063 &
065 only)
Electrical Specifications
Accuracy is ±(% reading number of digits) or otherwise specified, at 23
o
C ±
5
o
C & less than 75% R.H.
True RMS Model 062 & 065 ACV & ACA clamp-on accuracies are specified from 5% to
100% of range or otherwise specified. Maximum Crest Factor are as specified below,
and with frequency spectrums, besides fundamentals, fall within the meter specified AC
bandwidth for non-sinusoidal waveforms. Fundamentals are specified at 50Hz and
60Hz.
DC Voltage
RANGE
Accuracy
400.0 mV
0.3% + 3d
4.000V, 40.00V, 400.0V
0.5% + 3d
600V
1.0% + 4d
Input Impedance : 10M
, 30pF nominal (1000M
for 400.0mV range)
AC Voltage
RANGE
Accuracy
50Hz ~ 60Hz
4.000V, 40.00V, 400.0V
1.0% + 4d
60Hz ~ 500Hz
4.000V, 40.00V, 400.0V
1.5% + 4d
50Hz ~ 500Hz
600V
2.0% + 4d
Input Impedance: 10M
, 30pF nominal
True RMS Model 062 & 065 Crest Factor:
< 2 : 1 at full scale & < 4 : 1 at half scale