5.8  Phase Sequence Verification 

1.    Connect the red test lead to the “c” terminal, black 

test lead to “b” terminal and yellow test lead to “a” 

terminal 

2.  Set the multifunction selector to  “600V~/abc” 

range. 

3.  Connect yellow/black/red test  lead  to three-phase 

contact point, If the phase  sequence  indicator 

lights on, indicating the connection is in  correct 

phase sequence. 

4.  If the phase  sequence indicator  lights off, please 

swap  black (b) and red (c) test  leads, if the 

indicator  remains  off which means that there is a 

missing phase or wrong connection, please use 

Live Wire Verification function to verify each phase 

again. 

   

 

5.9  Live Wire Verification 

(BS1901/1W)

 

1.  Connect the red test lead to the “ ” terminal, black test lead to “COM” terminal and hold 

the black test lead (please note: 

do not touch the test pin and keep your hands away 

from the finger guard of the test lead!

2.  Set the multifunction selector to “600V~/abc” range. 

3.  Use the red test lead to contact the live wire to be tested.  When there is a live wire, the 

Live Wire Verification lights will turn on. 

 

5.10  Continuity test 

1.  Connect the black test lead to the “COM” terminal and red test lead to the “ ” terminal. 

2.  Set the multifunction selector to    (same as 

  ) position. 

3.  Connect  the test leads across the circuit to be tested, if the resistance less than 30

Ω, 

buzzer will be activated. 

!

CAUTION: Max. input overload: 250V rms < 20sec. 

 

5.11  Battery test 

(BS1901/01W)

 

1.  Connect the black test lead to the “COM” 

terminal and red test lead to the “batt.” 

terminal. 

2.  Set the multifunction selector to battery 

“1.5V or 12V” range 

3.  To measure the battery voltage without 

loading (V

1

), connect the test leads to 

the battery, red test lead to the battery 

cathode (+), back test lead to the battery 

anode (-).  If the display shows ‘

’ which 

means the red test lead is connected to 

the anode (-). 

4.  To measure the battery voltage  with 

loading (V

2

), press the “HOLD/batt.” key.  

The meter presets the internal loading R

=900

 for 12V and R

=150

 for 1.5V. 

5.  To find the internal impedance (R

i

) of the battery, the following formula can be used. 

R

i

 

Voltage without loading - Voltage with loading 

Internal loading 

 

(V

- V

2

)×R

0

 

V

2

 

For example:    V

1

=1.550V; V

2

=1.450V; R

0

 =150

 for 1.5V battery measurement 

 

 

 

 

5.12  Square waveform output

 (BS1901W)

 

1) Connect the black test lead to the “COM” socket and red test lead to the “

mA

” socket. 

2) Set the function selector to “

OUT

“position. 

3) connect test leads across device under test. 

 

 

 

5.13  Measuring AC current with an add-on clamp 

(optional) 

1.  Connect an add-on clamp anode (-)  to  “COM” 

terminal and cathode (+) to 

 

terminal 

2.  Clamp the tested wire with the clamp head and the 

AC current value will be shown on the display. 

 

 

 

 

 

 

6.  MAINTENANCE 

CAUTION 

BEFORE ATTEMPTING BATTERY REMOVAL OR REPLACEMENT, DISCONNECT 

TEST LEADS OR PROBES FROM ANY ENERGISED CIRCUITS TO AVOID SHOCK 

HAZARD. 

 

6.1  Fitting and replacing the battery   

1.  Ensure that the instrument is not connected to any external circuit, set the Multifunction 

selector to OFF position and remove the test leads from the terminals. 

2.  Remove the screw of the battery compartment on the bottom of the back case. 

3.  Replace the spent battery with the same type and rating. 

4.  Reinstate the battery compartment, tighten and securing screw. 

 

 

6.2  Cleaning 

Periodically wipe the case with a soft damp cloth and mild household cleanser.    Do not use 

abrasives or solvents.  Ensure that no water gets inside the equipment to prevent possible 

shorts and damage. 

 

7.  ACCESSORIES 

 

The accessories contained inside the packaging are the following: 

 

Pair of test leads 

 

English instruction manual 

 

3 Phases 3rd test lead with clip 

 

K-Type Thermocouple (for BS1903 only) 

 

10A Current Adaptor (for BS1903 only) 

 

1.5V AAA battery x 3   

 

Certificate of test 

Optional: 10A DC current Adaptor 3700- 10A 

         AC current add-on clamp BS- 06   

 

8.  SERVICE

 

8.1  Warranty Conditions 

 

This meter is guaranteed against any material fault or manufacturer’s defect, in accordance 

with the general conditions of sale. During the warranty period (one year), faulty parts may be 

replaced, with the manufacturer reserving the right to decide either to repair or replace the 

product. 

 

In the event of returning the meter to the after-sales service or to a regional branch, the 

outward transport is payable by the customer. The delivery must be agreed in advance with 

consignee. 

 

For delivery indicate, by means of an enclosed note, as clear as possible, the reasons for 

returning it. Use only the original packing. 

 

Any damage caused by shipment using NOT the original packaging will be charged in any 

case to the consignor. 

 

The manufacturer will not be responsible for any damage to persons or things. 

 

The warranty does not apply to the following cases: 

 

Accessories and battery are not included in warranty. 

 

Repairs following unsuitable use of the meter or by combining the latter with incompatible 

meter or accessories. 

 

Repairs resulting from incorrect shipping. 

 

Repairs resulting from servicing carried out by a person not approved by the company. 

 

Modifications to the meter without explicit authorisation from our technical department. 

 

Adaptation to a particular application not provided for by the definition of the meter or by 

the instruction manual. 

 

The contents of this manual may not be reproduced in any form whatsoever without the 

manufacturer’s authorization. 

 

Our products are patented. The logotypes are registered. We reserve the right to 

modify specifications and prices as part of technological developments which might 

be necessary. 

 

8.2  Service 

If the meter should not work properly, before contacting the DEALER OR THE SERVICE 

CENTRE, check the battery condition, the test leads, etc., Change them if necessary. 

If the meter still does not work, check if your operating procedure agrees with the description 

in this manual. 

 

In the event of returning the meter, it must be re-sent to the after-sales service (at address or 

to a regional branch), the outward transport is payable by the customer. The delivery must be 

agreed in advance with consignee. 

 

For delivery indicate the reasons for returning it. (By means of an enclosed note, as clear as 

possible).    Use only the original packing. 

 

Any damage caused by delivery with NO original packaging will be charged in any case to the 

consignor. 

 

FOR TECHNICAL ASSISTANCE, 

PLEASE CONTACT:   

 

 

 

 

 

R

i

  = 

(1.550 – 1.450)×150 

  10.34

 

1.450 

1901/1901W/1902/1903 

REV1/DEC 06 

!

Remarks: Due to our policy of continual product development, we reserve the 

right to amend the specifications of the mentioned products without notice.

 