3.4 - Probe cleaning system (option)
M2007/3en
TCT-2301 Installation instructions
35
©BTG 2004
3.4 Probe cleaning system (option)
This section describes how to install the PCD-1000 probe cleaning de-
vice. The system can be delivered as a separate unit for rebuilding exist-
ing units or together with a complete TCT-2301 transmitter. If the
cleaning system is ordered with the TCT-2301 the probe will be mounted
in the cleaning device at the BTG factory. The installation method is de-
scribed for both cases.
3.4.1
Dimensions and installation clearance
Ensure that there is sufficient clearance to remove the probe and cleaning
device from the valve assembly.
The installed device is approx. 700 mm (28 in) long. You will need at
least 300 mm (12 in) to remove the device from the valve arrangement.
The cleaning device weighs approx. 9.5 kg (21 lbs) including the TCT-
2301 probe.
Fig 27 PCD-1000 probe
cleaning device
140
5,5"
~550
22"
Min 100
4"
700
28"
Min 300
12"
92
3,6”
133
5,3"
Fig 28 PCD-1000 dimensions