4.4 - Other functions
M2007/3en
TCT-2301 Operating instructions
59
©BTG 2004
4.4.4
Manual loading of probe signal settings
This function is located under
,
.
It is a special function which can be useful in certain cases.
Procedure:
1. Go to MAIN MENU option
,
.
2. Step between values
and load the desired probe signal set-
ting by pressing
. Exit by pressing the
key.
3. Press
,
to send calibration data to the transmitter.
4-20mA MODEMPRINTER
PWR
EJECT
ON/OFF
MEMORY CARD
SPC-1000
Communication Terminal
Take
sample
menu
Load
lab.
cons.
Calc.
new
calibr.
Special
func.
Menu
Action:
1. Do configure before calibration.
2. Push <take sample> and collect sample
3. After lab. evaluation push <Load lab.
cons.>and enter your lab cons. value.
3. Push <Calc. new calibr.> to calculate
the new calibration.
<Special func.>: Set or adjust sample
values and lab values manually
Cali-
brate
Special
func.
Cali-
brate
Special
func.
SPC-1000
Communication Terminal
4-20mA MODEMPRINTER
PWR
EJECT
ON/OFF
MEMORY CARD
SPC-1000
Communication Terminal
No.ProbeLab.-Active
signal cons.
11.52%1.66%ON
21.77%1.86%ON
31.30%1.34%ON
41.09%1.10%ON
50.00%0.00%OFF
60.00%0.00%OFF
70.00%0.00%OFF
80.00%0.00%OFF
90.00%0.00%OFF
k0 = 0.1366
k1 = 0.8613
Menu
Edit
row
1
Load
lab.
cons.
Calibr.
curve
Load
probe
signal
ON/OFF
Edit
row
x
Load
probe
signal
Menu
Send
data
Send
change