4.6 - Electronics box display menu
70
TCT-2301 Operating instructions
M2007/3en
©BTG 2004
4.6.2.1
How to change settings and collect samples
Result display
1.30
%
TCONS
When the result display is shown, samples can be
collected for calibration purposes.
Press the sample
and collect a sample from the pipe
for lab evaluation.
NO:5
*
The display indicates the position in the calibration
table where the sample will be stored. The first empty
position in the table is selected.
Mark the lab sample with this number. Value X will be
counted down for each signal stored. An average will then
be calculated.
NO:5
1.30% OK?
Save the sample by pressing enter
, discard the
sample by pressing
or
.
SAMPLE SAVED
Enter
.
NOT SAVED
or
.
SAMPLE TAB FULL
If all nine samples in the calibration table are already
used, select the sample to be replaced with a new probe
signal value.
9
Press enter
and select which calibration sample to
replace using the
or
buttons. Verify with
enter
. Press
or
to return to result display
without replacing any calibration sample.
NO:9
*
The display indicates the position, take out the lab sample
at the same time.
NO:9
1.35% OK?
Enter
to save the value,
or
to
discard.