4.6 - Electronics box display menu
M2007/3en
TCT-2301 Operating instructions
71
©BTG 2004
!
Always mark the lab sample with the number indicated on the
display. The handheld terminal SPC-1000 must be used if you are
making a complete calibration.
Range setting
RANGE 1
Press enter
to select another range.
1
Press
or
to specify range. Verify by pressing
enter. Note! The transmitter is still operating on the range
set by the external switch.
Span setting
SPAN
Press enter
to change span.
0
.50-2.00 %
The first number is blinking, press
or
to change
the number, press enter
when ready and go to next
number and repeat the procedure.
1.00-2.
50
%
When the last number is blinking, press
or
to
change and enter
when ready.
SPAN
The result will be stored in the transmitter and the display
will return to the menu.
Offset adjustment
OFFSET
Press the sample
in this position and collect a
sample from the pipe for lab evaluation.
SAMPLE
*
1.77%
OK?
Press enter
to save the value, press
or
to
discard the value.