18
IMPORTANT!!!
1.You must press the C key to distinguish the control solution
test from the blood glucose test. Do not perform the blood
test in Control Solution mode.
(
If you perform the blood
test in Control Solution mode, the test result will not be
stored in the memory
.)
2. Use only
SuperCheck Plus
Control
Solution
with
SuperCheck Plus
Glucose
Meter
.
3.
SuperCheck Plus
Control Solutions are used to check
that the meter and the test strips are working together as a
system and that you are performing the test correctly.
4.
SuperCheck Plus
Control Solution contains a known
amount of glucose that reacts with test strips. The Medium
and High level Control Solution are intended to check the
monitoring system in different measurement ranges.
5.Shake the vial, discard the first drop of control solution, and
wipe off the dispenser tip to ensure a good sample and an
accurate result.
6.Use only for 3 months after first opening. Record the open
date on the Control Solution vial. Discard after 3 months.
7. Compare your Control Solution test results with the expected
range printed on the test strip vial label. If your glucose
control results fall outside the expected range; repeat the
test. Results that repeatedly fall outside the expected
range may indicate
:
(1)
You may not be doing the test correctly; repeat the test, by shaking
the control solution vial well and carefully following instructions.
Please make sure the test is done within the temperature range.
(
2
)
The Control Solution is expired or contaminated
.
(
3
)
The test strips are damaged or outdated.
(4)
You may have applied the Control Solution before the blood drop
appears on the screen. This will cause incorrect glucose measurement.
(
5
)
Meter malfunction
.