3 | Product description
Büchi Labortechnik AG
26/88
Operation Manual UniversalExtractor E-800
Specification
UniversalExtractor
E-800
Weight
(with glassware Universal)
52.6 kg
Total heating power
(rated)
1680 W
Total Heating power
(maximum)
1680 W
Hose connection
6 / 9 mm
Allowed water pressure
( nominal value)
6 bar
Allowed water pressure
(maximum)
8 bar
Minimum water flow
100 mL/min
Inlet cooling medium temperature
25 °C below the boiling
point of the solvent
Number of extraction positions
6
Solvent tank volume
2 L
Allowed inert gas pressure
(maximum)
3 bar
Max. filling level
(Extraction glass chamber Universal)
190 mL
Max. filling level
(Extraction glass chamber Universal LSV)
315 mL
Max. working volumes
(Beaker)
175 mL
Max. working volumes
(Beaker LSV)
320 mL
Language
DE, EN, IT, ES, FR, JA,
CN, PL, RU
Method storage
(Pro interface)
40 methods
3.5.2
Ambient conditions
For indoor use only.
Max. altitude above sea level
2000 m
Ambient temperature
5‒40 °C
Maximum relative humidity
80% for temperatures up to 31 °C
decreasing linearly to 50 % relative hu-
midity at 40 °C
Storage temperature
max. 45 °C