6 | Operating the interface pro
Büchi Labortechnik AG
50/88
Operation Manual UniversalExtractor E-800
Submenu Service
WARNING
Incorrect use of the actor settings.
Incorrect use of the actors settings can cause instrument damages and injuries.
u
Carry out the actor setting as described.
Action
Option
Explanation
[Sensors]
View
All sensor signals can be observed:
Flow senor
Level sensors
Analyte protection
Protection shield
Valves
Lift current
Line voltage
Line frequency
[Actors]
Open / close
The following settings can be
changed:
Valves:
Soxhlet valves See Draining solvent
from Extraction glass chamber
Soxhlet
Tank valves See Chapter 8.3
"Draining solvent from the
instrument", page 74
Cooling water valve On/Off
Gas valve On/Off (option)
Heater On/Off
Pump (solvent tank pump) On/Off
Lift current
u
Tap up or down.
ð
Display shows the actual current
consumption.
Beeper On/Off
Chiller On/Off (option)
Submenu System Information
The submenu
[System information]
contains details of the connected devices and
information on network connection diagnosis.